Prevalence and Risk Factors for Weaning Failure From Venovenous Extracorporeal Membrane Oxygenation in Patients With Severe Acute Respiratory Insufficiency.
Crit Care Med
; 52(1): 54-67, 2024 01 01.
Article
en En
| MEDLINE
| ID: mdl-37665263
ABSTRACT
OBJECTIVE:
Analysis of the prevalence and risk factors for weaning failure from venovenous extracorporeal membrane oxygenation (VV-ECMO) in patients with severe acute respiratory insufficiency.DESIGN:
Single-center retrospective observational study.SETTING:
Sixteen beds medical ICU at the University Hospital Regensburg. PATIENTS Two hundred twenty-seven patients with severe acute respiratory insufficiency requiring VV-ECMO support between October 2011 and December 2017.INTERVENTIONS:
None. MEASUREMENTS AND MAINRESULTS:
Patients meeting our ECMO weaning criteria (Sp o2 ≥ 90% with F io2 ≤ 0.4 or Pa o2 /F io2 > 150 mm Hg, pH = 7.35-7.45, positive end-expiratory pressure ≤ 10 cm H 2 O, driving pressure < 15 cm H 2 O, respiratory rate < 30/min, tidal volume > 5 mL/kg, ECMO bloodflow ≈ 1. 5 L/min, sweep gas flow ≈ 1 L/min, heart rate < 120/min, systolic blood pressure 90-160 mm Hg, norepinephrine < 0.2 µg/[kg*min]) underwent an ECMO weaning trial (EWT) with pausing sweep gas flow. Arterial blood gas analysis, respiratory and ventilator parameters were recorded prior, during, and after EWTs. Baseline data, including demographics, vitals, respiratory, ventilator, and laboratory parameters were recorded at the time of cannulation. One hundred seventy-nine of 227 (79%) patients were successfully decannulated. Ten patients (4%) underwent prolonged weaning of at least three failed EWTs before successful decannulation. The respiratory rate (19/min vs 16/min, p = 0.002) and Pa co2 (44 mm Hg vs 40 mm Hg, p = 0.003) were higher before failed than successful EWTs. Both parameters were risk factors for ECMO weaning failure (Pa co2 odds ratio [OR] 1.05; 95% CI, 1.001-1.10; p = 0.045; respiratory rate OR 1.10; 95% CI, 1.04-1.15; p < 0.001) in multivariable analysis. The rapid shallow breathing index [42 (1/L*min), vs 35 (1/L*min), p = 0.052) was higher before failed than successful EWTs. The decline of Sa o2 and Pa o2 /F io2 during EWTs was higher in failed than successful trials.CONCLUSIONS:
Seventy-nine percent of patients were successfully decannulated with only 4% needing prolonged ECMO weaning. Before EWT only parameters of impaired ventilation (insufficient decarboxylation, higher respiratory rate) but not of oxygenation were predictive for weaning failure, whereas during EWT-impaired oxygenation was associated with weaning failure.
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Síndrome de Dificultad Respiratoria
/
Insuficiencia Respiratoria
/
Oxigenación por Membrana Extracorpórea
Tipo de estudio:
Etiology_studies
/
Observational_studies
/
Prevalence_studies
/
Prognostic_studies
/
Risk_factors_studies
Límite:
Humans
Idioma:
En
Revista:
Crit Care Med
Año:
2024
Tipo del documento:
Article
País de afiliación:
Alemania