RESUMEN

Flow diverters (FDs) are utilized for a wide range of aneurysms, but show safety issues such as adverse interactions with static magnetic fields (displacement force and torque) and radiofrequency-induced heating during magnetic resonance imaging (MRI). The present study aimed to assess these adverse interactions in a 7-tesla (7T) static magnetic field and radiofrequency-induced heating during a 7T MRI for two types of FD. Displacement force and magnetically induced torque were assessed using the deflection angle method and low friction surface method, respectively. To assess heating, each FD was set in a phantom filled with gelled-saline mixed with polyacrylic acid and underwent a 7T MRI using a three-dimensional fast spin echo method. Displacement force and magnetically induced torque in the 7T static magnetic field were undetectable, and radiofrequency-induced heating during 7T MRI remained ≤ 0.6 °C for both types of FD, suggesting that magnetic field interactions and heating on FDs during a 7T MRI are acceptable from a safety perspective.


Asunto(s)
Aneurisma Intracraneal , Humanos , Aneurisma Intracraneal/diagnóstico por imagen , Calefacción , Campos Magnéticos , Imagen por Resonancia Magnética/efectos adversos , Imagen por Resonancia Magnética/métodos