Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros

Banco de datos
Tipo del documento
Asunto de la revista
País de afiliación
Intervalo de año de publicación
1.
Curr Cardiol Rep ; 24(8): 947-958, 2022 08.
Artículo en Inglés | MEDLINE | ID: mdl-35639275

RESUMEN

PURPOSE OF REVIEW: While the subcutaneous (S-) implantable cardioverter-defibrillator (ICDs) is an alternative to the transvenous (TV-) ICD in many patients, optimal use remains unclear. In this review, we summarize recent clinically relevant data on sensing algorithms, inappropriate shocks, defibrillation testing, and battery and electrode failures. RECENT FINDINGS: Changes in sensing algorithms and S-ICD programming have significantly decreased inappropriate shock rates. Avoiding fat below the S-ICD coil and can is key for reducing the defibrillation threshold. While S-ICD battery and electrode failures have resulted in recalls, system components remain commercially available since failure rates are low and no other similar devices are available. The S-ICD is a good alternative to the TV-ICD for many patients, and particularly in light of recently developed device algorithms and improvements in implant technique. Future research will need to better understand: the impact of S-ICD electrode and battery failures and the potential for integrating leadless pacing into a modular S-ICD platform.


Asunto(s)
Desfibriladores Implantables , Muerte Súbita Cardíaca , Humanos , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA