Artículo en Zh | WPRIM | ID: wpr-910353

RESUMEN

Objective:To discuss the problems existing in the implementation of the Chinese health industry standard WS 523-2019 by testing SPECT device.Methods:Under the WS 523-2019 standards, a total of 10 SPECT devices were tested with regard to their SPECT reconstructed spatial resolution (SRSR), system planar sensitivity (SPS), system spatial resolution (SSR), whole body system spatial resolution (WSSR), intrinsic uniformity (IU), intrinsic count rate performance (ICR), intrinsic spatial resolution (ISR) and intrinsic spatial linearity(ISL).Result:Under the requirements of WS 523-2019 standards for qualified limits, there are 3 devices with ISL unqualified and the rest of the performances qualified. The new standards basically can meet the clinical requirements and reflected the overall performance of SPECT.Conclusions:The distance between the radiation source and the surface of the detector has great influence on the spatial resolution.In the measurement of ISL, there must be a lead grid separately in the x and y directions. The lead grids with the parallel slits shall be positioned on the detector with the center slit centered on the detector. It is suggested to add rotation center in the new standards.