Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy.
Shirato, Hiroki; Suzuki, Keishiro; Sharp, Gregory C; Fujita, Katsuhisa; Onimaru, Rikiya; Fujino, Masaharu; Kato, Norio; Osaka, Yasuhiro; Kinoshita, Rumiko; Taguchi, Hiroshi; Onodera, Shunsuke; Miyasaka, Kazuo.
Afiliación
  • Shirato H; Department of Radiology, Hokkaido University School of Medicine, Sapporo, Japan. hshirato@radi.med.hokudai.ac.jp
Int J Radiat Oncol Biol Phys ; 64(4): 1229-36, 2006 Mar 15.
Article en En | MEDLINE | ID: mdl-16504762
ABSTRACT

BACKGROUND:

To reduce the uncertainty of registration for lung tumors, we have developed a four-dimensional (4D) setup system using a real-time tumor-tracking radiotherapy system. METHODS AND MATERIALS During treatment planning and daily setup in the treatment room, the trajectory of the internal fiducial marker was recorded for 1 to 2 min at the rate of 30 times per second by the real-time tumor-tracking radiotherapy system. To maximize gating efficiency, the patient's position on the treatment couch was adjusted using the 4D setup system with fine on-line remote control of the treatment couch.

RESULTS:

The trajectory of the marker detected in the 4D setup system was well visualized and used for daily setup. Various degrees of interfractional and intrafractional changes in the absolute amplitude and speed of the internal marker were detected. Readjustments were necessary during each treatment session, prompted by baseline shifting of the tumor position.

CONCLUSION:

The 4D setup system was shown to be useful for reducing the uncertainty of tumor motion and for increasing the efficiency of gated irradiation. Considering the interfractional and intrafractional changes in speed and amplitude detected in this study, intercepting radiotherapy is the safe and cost-effective method for 4D radiotherapy using real-time tracking technology.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Tomografía Computarizada por Rayos X / Radioterapia Conformacional / Neoplasias Pulmonares / Movimiento Límite: Humans Idioma: En Revista: Int J Radiat Oncol Biol Phys Año: 2006 Tipo del documento: Article País de afiliación: Japón
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Tomografía Computarizada por Rayos X / Radioterapia Conformacional / Neoplasias Pulmonares / Movimiento Límite: Humans Idioma: En Revista: Int J Radiat Oncol Biol Phys Año: 2006 Tipo del documento: Article País de afiliación: Japón