Determination of butylated hydroxytoluene in food samples by high-performance liquid chromatography with ultraviolet detection and gas chromatography/mass spectrometry.
J AOAC Int
; 90(1): 277-83, 2007.
Article
en En
| MEDLINE
| ID: mdl-17373461
A reversed-phase high-performance liquid chromatography (RP-HPLC) method was developed and compared with a gas chromatography/mass spectrometry (GC/MS) method for determining butylated hydroxytoluene (BHT) in foodstuffs as a result of migration from plastic packaging. Similar extraction procedures were used in both methods. BHT was quantitated using an external standard in the HPLC method and an internal standard in the GC/MS method. Both methods presented good linearity (r(2) > or = 0.9917) and low detection limits. Recoveries obtained with the HPLC method (chicken meat, 95.8%, and Gouda cheese, 83.9%) were better than with the GC/MS method (chicken meat, 85.6%, and Gouda cheese, 71.3%).
Buscar en Google
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Hidroxitolueno Butilado
/
Cromatografía Líquida de Alta Presión
/
Análisis de los Alimentos
Tipo de estudio:
Diagnostic_studies
Límite:
Animals
Idioma:
En
Revista:
J AOAC Int
Año:
2007
Tipo del documento:
Article
País de afiliación:
España