Safety and efficacy of high-intensity focused ultrasound atop coronary arteries during epicardial catheter ablation.
Koruth, Jacob S; Dukkipati, Srinivas; Carrillo, Roger G; Coffey, James; Teng, Jennifer; Eby, Thomas B; Reddy, Vivek Y; D'Avila, Andre.
Afiliación
  • Koruth JS; Helmsley Cardiac Arrhythmia Service, Mount Sinai Medical Center, New York, New York, USA.
J Cardiovasc Electrophysiol ; 22(11): 1274-80, 2011 Nov.
Article en En | MEDLINE | ID: mdl-21676047
ABSTRACT

BACKGROUND:

Coronary arterial injury continues to be a limitation of epicardial catheter ablation using currently available energy sources. Application of high intensity focused ultrasound (HIFU) energy may avoid such injury due to its theoretical ability to focus energy beyond the ablation element and create lesions at depth.

OBJECTIVE:

This study evaluated the safety of HIFU applications delivered directly over the left anterior descending (LAD) artery in an open-chest swine model.

METHODS:

Ten swine underwent median sternotomy. A prototype HIFU probe was placed atop the LAD. Forty-three therapies along the LAD (60-seconds/6 watt) were analyzed. Three, 3, and 4 swine were studied at 2, 4, and 8 weeks and subsequently sacrificed. Lesions were scored (0-4) depending on the percent circumferential involvement of arteries.

RESULTS:

Lesion area increased minimally from 54.5 ± 18.0 mm(2) at 2 weeks to 56.9 ± 20.6 mm(2) at 8 weeks, and depth increased moderately from 13.2 ± 2.5 mm to 15.5 ± 3.4 mm. At 2, 4, and 8 weeks, the mean injury score of the LAD was 0.8 ± 0.3, 1.5 ± 0.9, and 2.0 ± 0.7. No/minimal arterial injury was seen in 64% of all sections. However, a progressive increase in injury resulted in 89% of all sections showing any injury at 8 weeks. One animal developed occlusion of the distal LAD.

CONCLUSIONS:

HIFU has the potential to create deep ventricular lesions with relative sparing of the LAD. The incremental arterial damage noted over time warrants further evaluation to support the viability of focusing ultrasound energy beyond vulnerable critical structures to ablate deeper targets.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Pericardio / Ablación por Catéter / Ultrasonido Enfocado de Alta Intensidad de Ablación / Lesiones del Sistema Vascular / Lesiones Cardíacas Tipo de estudio: Etiology_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Pericardio / Ablación por Catéter / Ultrasonido Enfocado de Alta Intensidad de Ablación / Lesiones del Sistema Vascular / Lesiones Cardíacas Tipo de estudio: Etiology_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos