First-in-man evaluation of the novel balloon delivery system STENTYS Xposition S for the self-apposing coronary artery stent: impact on longitudinal geographic miss during stenting.
EuroIntervention
; 11(12): 1341-5, 2016 Mar.
Article
en En
| MEDLINE
| ID: mdl-25983027
ABSTRACT
AIMS:
A novel balloon delivery system (BDS) for the self-apposing STENTYS sirolimus-eluting stent (SES) has been developed for highly precise longitudinal stent positioning and deployment. The aim of this first-in-man study is to report the quantitative coronary analysis (QCA) angiography and optical coherence tomography (OCT) results as well as the 30-day clinical outcomes of the STENTYS Xposition S SES. METHODS ANDRESULTS:
We included 25 patients (mean age 66.1±10.7 years) with stable coronary artery disease (24%) or acute coronary syndrome (including STEMI in 40%). The device was successfully placed at the intended site in all cases (100%), without procedural complications. Longitudinal geographic miss (entire lesion length [on QCA] not completely covered by the stent) was not observed. Pre-procedural MLD on QCA was 1.30±0.74 mm and post-procedural MLD was 2.74±0.44 mm, p<0.001 (acute gain 1.44±0.70 mm). OCT analyses showed a low percentage of malapposed stent struts directly post stent placement (2.4%), which further decreased after post-dilatation (0.6%, p=0.013), while mean stent area increased (from 9.7 mm2 to 10.5 mm2, p<0.001). At 30-day clinical follow-up, one (4%) major adverse cardiac event (MACE) was observed. One acute stent thrombosis (ST) occurred immediately post procedure in a STEMI patient which was related to inadequate medication therapy.CONCLUSIONS:
This first-in-man study demonstrated that the use of the novel STENTYS Xposition S balloon delivery system is feasible with a high technical success rate without longitudinal geographical miss. Stent strut malapposition rate directly after STENTYS placement was low and improved further after post-dilatation.
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Enfermedad de la Arteria Coronaria
/
Cateterismo Cardíaco
/
Angioplastia Coronaria con Balón
/
Síndrome Coronario Agudo
/
Stents Liberadores de Fármacos
/
Catéteres Cardíacos
/
Infarto del Miocardio con Elevación del ST
Tipo de estudio:
Clinical_trials
/
Etiology_studies
/
Observational_studies
Límite:
Aged
/
Female
/
Humans
/
Male
/
Middle aged
País/Región como asunto:
Europa
Idioma:
En
Revista:
EuroIntervention
Asunto de la revista:
ANGIOLOGIA
/
CARDIOLOGIA
/
TERAPEUTICA
Año:
2016
Tipo del documento:
Article
País de afiliación:
Países Bajos