Normal ranges and test-retest reproducibility of flow and velocity parameters in intracranial arteries measured with phase-contrast magnetic resonance imaging.
Correia de Verdier, Maria; Wikström, Johan.
Afiliación
  • Correia de Verdier M; Department of Radiology, Uppsala University, Uppsala University Hospital, 751 85, Uppsala, Sweden. mariacdev@hotmail.com.
  • Wikström J; Department of Radiology, Uppsala University, Uppsala University Hospital, 751 85, Uppsala, Sweden.
Neuroradiology ; 58(5): 521-31, 2016 May.
Article en En | MEDLINE | ID: mdl-26882908
ABSTRACT

INTRODUCTION:

The purpose of the present study was to investigate normal ranges and test-retest reproducibility of phase-contrast MRI (PC-MRI)-measured flow and velocity parameters in intracranial arteries.

METHODS:

Highest flow (HF), lowest flow (LF), peak systolic velocity (PSV), and end diastolic velocity (EDV) were measured at two dates in the anterior (ACA), middle (MCA), and posterior (PCA) cerebral arteries of 30 healthy volunteers using two-dimensional PC-MRI at 3 T. Least detectable difference (LDD) was calculated.

RESULTS:

In the left ACA, HF was (mean (range, LDD)) 126 ml/min (36-312, 59 %), LF 61 ml/min (0-156, 101 %), PSV 64 cm/s (32-141, 67 %), and EDV 35 cm/s (18-55, 42 %); in the right ACA, HF was 154 ml/min (42-246, 49 %), LF 77 ml/min (0-156, 131 %), PSV 75 cm/s (26-161, 82 %), and EDV 39 cm/s (7-59, 67 %). In the left MCA, HF was 235 ml/min (126-372, 35 %), LF 116 ml/min (42-186, 48 %), PSV 90 cm/s (55-183, 39 %), and EDV 46 cm/s (20-66, 28 %); in the right MCA, HF was 238 ml/min (162-342, 44 %), LF 120 ml/min (72-216, 48 %), PSV 88 cm/s (55-141, 35 %), and EDV 45 cm/s (26-67, 23 %). In the left PCA, HF was 108 ml/min (42-168, 54 %), LF 53 ml/min (18-108, 64 %), PSV 50 cm/s (24-77, 63 %), and EDV 28 cm/s (14-40, 45 %); in the right PCA, HF was 98 ml/min (30-162, 49 %), LF 49 ml/min (12-84, 55 %), PSV 47 cm/s (27-88, 59 %), and EDV 27 cm/s (16-41, 45 %).

CONCLUSION:

PC-MRI-measured flow and velocity parameters in the main intracranial arteries have large normal ranges. Reproducibility is highest in MCA.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Velocidad del Flujo Sanguíneo / Arterias Cerebrales / Circulación Cerebrovascular / Angiografía por Resonancia Magnética Tipo de estudio: Diagnostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Neuroradiology Año: 2016 Tipo del documento: Article País de afiliación: Suecia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Velocidad del Flujo Sanguíneo / Arterias Cerebrales / Circulación Cerebrovascular / Angiografía por Resonancia Magnética Tipo de estudio: Diagnostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Neuroradiology Año: 2016 Tipo del documento: Article País de afiliación: Suecia