Multi-angle fluorometer technique for the determination of absorption and scattering coefficients of subwavelength nanoparticles.
Opt Express
; 24(15): 17090-102, 2016 Jul 25.
Article
en En
| MEDLINE
| ID: mdl-27464160
ABSTRACT
A thorough analysis of the resonance light scattering (RLS) technique for quantitative scattering measurements of subwavelength nanoparticles is reported. The systematic error associated with using a measurement at a single angle to represent all of the scattered light is investigated. In-depth analysis of the reference material was performed to identify and minimize the error associated with the reference material. Semiconductor ZnO nanobullets and spherical Au nanoparticles of various sizes were used to verify the approach. A simple and inexpensive modification to standard fluorometers is demonstrated using a glass prism allowing scattering measurements in the slightly forward and backwards directions. This allows quantification of the systematic error associated with RLS which is consistently overlooked.