Variable impact of CSF flow suppression on quantitative 3.0T intracranial vessel wall measurements.
Cogswell, Petrice M; Siero, Jeroen C W; Lants, Sarah K; Waddle, Spencer; Davis, L Taylor; Gilbert, Guillaume; Hendrikse, Jeroen; Donahue, Manus J.
Afiliación
  • Cogswell PM; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Siero JCW; Department of Radiology, University Medical Center Utrecht, Utrecht, The Netherlands.
  • Lants SK; Spinoza Center for Neuroimaging, Amsterdam, The Netherlands.
  • Waddle S; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Davis LT; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Gilbert G; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Hendrikse J; MR Clinical Science, Philips Healthcare Canada, Markham, Ontario, Canada.
  • Donahue MJ; Department of Radiology, University Medical Center Utrecht, Utrecht, The Netherlands.
J Magn Reson Imaging ; 48(4): 1120-1128, 2018 10.
Article en En | MEDLINE | ID: mdl-29603829
ABSTRACT

BACKGROUND:

Flow suppression techniques have been developed for intracranial (IC) vessel wall imaging (VWI) and optimized using simulations; however, simulation results may not translate in vivo.

PURPOSE:

To evaluate experimentally how IC vessel wall and lumen measurements change in identical subjects when evaluated using the most commonly available blood and cerebrospinal fluid (CSF) flow suppression modules and VWI sequences. STUDY TYPE Prospective. POPULATION/

SUBJECTS:

Healthy adults (n = 13; age = 37 ± 15 years) were enrolled. FIELD STRENGTH/SEQUENCE A 3.0T 3D T1 /proton density (PD)-weighted turbo-spin-echo (TSE) acquisition with post-readout anti-driven equilibrium module, with and without Delay-Alternating-with-Nutation-for-Tailored-Excitation (DANTE) was applied. DANTE flip angle (8-12°) and TSE refocusing angle (sweep = 40-120° or 50-120°) were varied. ASSESSMENT Basilar artery and internal carotid artery (ICA) wall thicknesses, CSF signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and signal ratio (SR) were assessed. Measurements were made by two readers (radiology resident and board-certified neuroradiologist). STATISTICAL TESTS A Wilcoxon signed-rank test was applied with corrected two-sided P < 0.05 required for significance (critical P = 0.008, 0.005, and 0.05 for SNR/CNR, SR, and wall thickness, respectively).

RESULTS:

A TSE pulse sweep = 40-120° and sweep = 50-120° provided similar (P = 0.55) CSF suppression. Addition of the DANTE preparation reduced CSF SNR from 17.4 to 6.7, thereby providing significant (P < 0.008) improvement in CSF suppression. The DANTE preparation also resulted in a significant (P < 0.008) reduction in vessel wall SNR, but variable vessel wall to CSF CNR improvement (P = 0.87). There was a trend for a difference in blood SNR with vs. without DANTE (P = 0.05). The outer vessel wall diameter and wall thickness values were lower (P < 0.05) with (basilar artery 4.45 mm, 0.81 mm, respectively) vs. without (basilar artery 4.88 mm, 0.97 mm, respectively) DANTE 8°. DATA

CONCLUSION:

IC VWI with TSE sweep = 40-120° and with DANTE flip angle = 8° provides the best CSF suppression and CNR of the approaches evaluated. However, improvements are heterogeneous, likely owing to intersubject vessel pulsatility and CSF flow variations, which can lead to variable flow suppression efficacy in these velocity-dependent modules. LEVEL OF EVIDENCE 2 Technical Efficacy Stage 1 J. Magn. Reson. Imaging 2018;481120-1128.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Encéfalo / Líquido Cefalorraquídeo / Angiografía por Resonancia Magnética Tipo de estudio: Observational_studies / Prognostic_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Encéfalo / Líquido Cefalorraquídeo / Angiografía por Resonancia Magnética Tipo de estudio: Observational_studies / Prognostic_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos