Your browser doesn't support javascript.
loading
Head-to-head comparison of a CZT-based all-purpose SPECT camera and a dedicated CZT cardiac device for myocardial perfusion and functional analysis.
Gimelli, Alessia; Liga, Riccardo; Bertasi, Michelangelo; Kusch, Annette; Marzullo, Paolo.
Afiliación
  • Gimelli A; Fondazione Toscana/CNR G. Monasterio, Via Moruzzi, 1, 56124, Pisa, Italy. gimelli@ftgm.it.
  • Liga R; Università di Pisa, Pisa, Italy.
  • Bertasi M; Fondazione Toscana/CNR G. Monasterio, Via Moruzzi, 1, 56124, Pisa, Italy.
  • Kusch A; Fondazione Toscana/CNR G. Monasterio, Via Moruzzi, 1, 56124, Pisa, Italy.
  • Marzullo P; Fondazione Toscana/CNR G. Monasterio, Via Moruzzi, 1, 56124, Pisa, Italy.
J Nucl Cardiol ; 28(4): 1323-1330, 2021 Aug.
Article en En | MEDLINE | ID: mdl-31385223
ABSTRACT

PURPOSE:

To compare the outputs of a novel all-purpose SPECT camera equipped with CZT detectors (Discovery NM/CT 670) with the state-of-the-art represented by a dedicated CZT (Alcyone, Discovery 530c) cardiac camera in patients submitted to myocardial perfusion imaging (MPI).

METHODS:

We included 19 patients that underwent sequential low-dose 99mTc-tetrofosmin (148-185 MBq during stress and 296-370 MBq at rest) MPI with Alcyone and Discovery 670 cameras. Quantitative (% tracer's uptake) and semi-quantitative analyses of perfusion data were performed for each scan. Moreover, major left ventricular (LV) functional and structural parameters were derived from each camera and compared.

RESULTS:

The two cameras showed excellent correlation for segmental myocardial % uptake at stress (R = 0.90; P < 0.001) and at rest (R = 0.88; P < 0.001) with narrow Bland-Altman limits of agreement. The level of diagnostic agreement of Discovery 670 and Alcyone cameras regarding perfusion analysis was excellent (Cohen's κ 0.85). Similarly, the two cameras showed excellent correlation in the evaluation of LV ejection fraction (R = 0.95), peak filling rate (R = 0.97), and mass (R = 0.98).

CONCLUSIONS:

Our preliminary results suggest that MPI with an all-purpose Discovery 670 CZT-SPECT camera is feasible, comparing well with the current state-of-the-art technology.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Telurio / Zinc / Cadmio / Cámaras gamma / Tomografía Computarizada de Emisión de Fotón Único / Función Ventricular Izquierda / Imagen de Perfusión Miocárdica / Cardiopatías Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Telurio / Zinc / Cadmio / Cámaras gamma / Tomografía Computarizada de Emisión de Fotón Único / Función Ventricular Izquierda / Imagen de Perfusión Miocárdica / Cardiopatías Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Italia