Your browser doesn't support javascript.
loading
Myocardial T1 and T2 quantification and water-fat separation using cardiac MR fingerprinting with rosette trajectories at 3T and 1.5T.
Liu, Yuchi; Hamilton, Jesse; Eck, Brendan; Griswold, Mark; Seiberlich, Nicole.
Afiliación
  • Liu Y; Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA.
  • Hamilton J; Department of Radiology, University of Michigan, Ann Arbor, MI, USA.
  • Eck B; Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA.
  • Griswold M; Department of Radiology, University of Michigan, Ann Arbor, MI, USA.
  • Seiberlich N; Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA.
Magn Reson Med ; 85(1): 103-119, 2021 01.
Article en En | MEDLINE | ID: mdl-32720408
ABSTRACT

PURPOSE:

This work aims to develop an approach for simultaneous water-fat separation and myocardial T1 and T2 quantification based on the cardiac MR fingerprinting (cMRF) framework with rosette trajectories at 3T and 1.5T.

METHODS:

Two 15-heartbeat cMRF sequences with different rosette trajectories designed for water-fat separation at 3T and 1.5T were implemented. Water T1 and T2 maps, water image, and fat image were generated with B0 inhomogeneity correction using a B0 map derived from the cMRF data themselves. The proposed water-fat separation rosette cMRF approach was validated in the International Society for Magnetic Resonance in Medicine/National Institute of Standards and Technology MRI system phantom and water/oil phantoms. It was also applied for myocardial tissue mapping of healthy subjects at both 3T and 1.5T.

RESULTS:

Water T1 and T2 values measured using rosette cMRF in the International Society for Magnetic Resonance in Medicine/National Institute of Standards and Technology phantom agreed well with the reference values. In the water/oil phantom, oil was well suppressed in the water images and vice versa. Rosette cMRF yielded comparable T1 but 2~3 ms higher T2 values in the myocardium of healthy subjects than the original spiral cMRF method. Epicardial fat deposition was also clearly shown in the fat images.

CONCLUSION:

Rosette cMRF provides fat images along with myocardial T1 and T2 maps with significant fat suppression. This technique may improve visualization of the anatomical structure of the heart by separating water and fat and could provide value in diagnosing cardiac diseases associated with fibrofatty infiltration or epicardial fat accumulation. It also paves the way toward comprehensive myocardial tissue characterization in a single scan.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Imagen por Resonancia Magnética / Agua / Corazón Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Imagen por Resonancia Magnética / Agua / Corazón Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: Estados Unidos