PERFORMANCE OF HANDHELD NAI(TL) SPECTROMETERS AS DOSIMETERS BY LABORATORY AND FIELD DOSE RATE MEASUREMENTS.
Clouvas, A; Xanthos, S; Boziari, A; Leontaris, F; Kaissas, I; Omirou, M.
Afiliación
  • Clouvas A; Department of Electrical and Computer Engineering, Aristotle University of Thessaloniki, Thessaloniki GR-54124, Greece.
  • Xanthos S; Department of Industrial Engineering and Management, International Hellenic University, Thessaloniki GR-57001, Greece.
  • Boziari A; Greek Atomic Energy Commission, Agia Paraskevi GR-15310, Greece.
  • Leontaris F; Department of Electrical and Computer Engineering, Aristotle University of Thessaloniki, Thessaloniki GR-54124, Greece.
  • Kaissas I; Greek Atomic Energy Commission, Agia Paraskevi GR-15310, Greece.
  • Omirou M; Department of Electrical and Computer Engineering, Aristotle University of Thessaloniki, Thessaloniki GR-54124, Greece.
Radiat Prot Dosimetry ; 194(4): 233-248, 2021 Aug 02.
Article en En | MEDLINE | ID: mdl-34260730
ABSTRACT
In the framework of the IAEA Coordinated Research Project (CRP) J02012 on 'Advancing Radiation Detection Equipment for Detecting Nuclear and Other Radioactive Material Out of Regulatory Control', the properties of two commercial instruments (1) InSpector 1000 analyzer (Canberra), with a 2″ × 2″ NaI(Tl) scintillator and (2) RIIDEYE M-G3 analyzer (Thermo Scientific), with a 3″ × 3″ NaI(Tl) scintillator, were evaluated as dosimeters by laboratory and field measurements. In the Ionizing Radiation Calibration Laboratory (IRCL) of the Greek Atomic Energy Commission, the NaI(Tl) spectrometers were tested in order to measure Ambient gamma Dose Equivalent Rate (ADER). The NaI(Tl) scintillators were irradiated in a homogeneous field with 662 keV photons with different ADER values from 0.17 to 100 µSv h-1 at 0° incidence (irradiation field perpendicular to the detector's front window) and at 90° incidence. For each irradiation, the measured ADER by the spectrometers and the 'true' ADER values (provided by the IRCL) were compared. In addition, the angular dependence (0-359°) of the ADER response of the spectrometers was studied with a 152Eu source placed at 1, 2 and 3 m from the spectrometers. The ADER dependence as function of the distance from the 152Eu source (at 0° incidence) measured by the two detectors was compared with the theoretical one. In the field studies, ADER was measured by the spectrometers at seven locations belonging to the Greek Early Warning System Network (which is based on Reuter-Stokes ionization chambers). These locations have different ADER values ranging from 20 to 120 nSv h-1. In these locations, gamma ADER were also deduced (1) by in situ gamma spectrometry measurements with portable Germanium HPGe detectors and (2) by the Reuter-Stokes ionization chambers (by subtraction of the cosmic radiation). Gamma dose measurements were also performed with the InSpector 1000 and RIIDEYE M-G3 detectors in 25 locations (beaches) of Northern Greece and compared with the ADER values deduced by sand sample analysis with gamma spectroscopy. Beaches with sand are good candidates for such type of measurements since they are commonly flat and in principle the natural radionuclides are homogenously distributed.