A transformer-coupled double-resonant probe for NMR imaging and spectroscopy.
Fitzsimmons, J R; Brooker, H R; Beck, B.
Afiliación
  • Fitzsimmons JR; Department of Radiology, University of Florida, Gainesville 32610.
Magn Reson Med ; 5(5): 471-7, 1987 Nov.
Article en En | MEDLINE | ID: mdl-3431409
ABSTRACT
A new type of double-resonant coil which takes advantage of the properties of a radio-frequency transformer is described. Two concentric loops are wound in close proximity to yield a high mutual inductance. The primary side is tuned to 34 MHz for 31P and the secondary side is tuned to 85 MHz for protons. A single lead allows program control over frequency without the need to rearrange cabling. Proton and phosphorus spectroscopy are made possible over essentially the same volume of interest without the need for repositioning the sample.