An evaluation of the IOLMaster 700 and its agreement with the IOLMaster v3 in children.
Ophthalmic Physiol Opt
; 42(1): 48-58, 2022 01.
Article
en En
| MEDLINE
| ID: mdl-34761427
PURPOSE: To evaluate the repeatability and reproducibility of the swept-source optical coherence tomographer Zeiss IOLMaster 700 and compare its outputs with those obtained using partial coherence interferometry (Zeiss IOLMaster v3) in a healthy, paediatric population. METHODS: This is a cross-sectional, observational study. Examiner 1 took two sets of biometric measurements (axial length [AL], mean corneal radius of curvature [Kmean ], anterior chamber depth [ACD] and lens thickness [LT]) using the IOLMaster 700, and one set of measurements (AL, Kmean and ACD) using the IOLMaster v3. Examiner 2 took one full set of measurements using the IOLMaster 700. Mean differences and 95% limits of agreement (LOA) were calculated, and Bland and Altman plots used to explore repeatability and reproducibility of the IOLMaster 700 alongside establishing its agreement with the IOLMaster v3. RESULTS: Mean participant age was 7.52 ± 0.58 years. Repeatability analyses demonstrated small mean differences and narrow 95% LOA for AL (0.001, -0.013 to 0.015 mm), Kmean (0.002, -0.020 to 0.024 mm), ACD (-0.003, -0.031 to 0.024 mm) and LT (0.001, -0.024 to 0.026 mm), respectively. Similarly, small mean differences and narrow 95% LOA established excellent reproducibility (AL 0.001, -0.016 to 0.018 mm; Kmean -0.001, -0.027 to 0.025 mm; ACD -0.010, -0.041 to 0.021 mm; LT 0.002, -0.016 to 0.020 mm). The IOLMaster 700 and IOLMaster v3 demonstrated good agreement with small mean differences and narrow 95% LOA (AL 0.009, -0.034 to 0.052 mm; Kmean 0.016, -0.013 to 0.044 mm; ACD 0.134, 0.055 to 0.212 mm). CONCLUSIONS: When used within a paediatric population, these data demonstrate the IOLMaster 700 to be highly repeatable and reproducible for measures of AL, Kmean , ACD and LT. There is excellent inter-instrument agreement between the IOLMaster 700 and IOLMaster v3 for measures of AL and Kmean . ACD measurements show weaker agreement. These data will be useful when considering reports from population-based studies of refractive error and clinical myopia research.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Longitud Axial del Ojo
/
Cámara Anterior
Tipo de estudio:
Observational_studies
/
Prevalence_studies
/
Risk_factors_studies
Límite:
Child
/
Humans
Idioma:
En
Revista:
Ophthalmic Physiol Opt
Año:
2022
Tipo del documento:
Article