Validation of the accuracy of contact force measurement by contemporary force-sensing ablation catheters.
Kueffer, Thomas; Haeberlin, Andreas; Knecht, Sven; Baldinger, Samuel H; Madaffari, Antonio; Seiler, Jens; Mühl, Aline; Tanner, Hildegard; Roten, Laurent; Reichlin, Tobias.
Afiliación
  • Kueffer T; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Haeberlin A; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Knecht S; ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland.
  • Baldinger SH; Department of Cardiology, University Hospital Basel, Basel, Switzerland.
  • Madaffari A; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Seiler J; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Mühl A; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Tanner H; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Roten L; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Reichlin T; Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
J Cardiovasc Electrophysiol ; 34(2): 292-299, 2023 02.
Article en En | MEDLINE | ID: mdl-36490307
ABSTRACT

INTRODUCTION:

Contact force sensing catheters are widely used for ablation of cardiac arrhythmias. They allow quantification of catheter-to-tissue contact, which is an important determinant for lesion formation and may reduce the risk of complications. The accuracy of these sensors may vary across the measurement range, catheter-to-tissue angle, and amongst manufacturers. We aim to compare the accuracy and reproducibility of four different force sensing ablation catheters.

METHODS:

A measurement setup containing a heated saline water bath with an integrated force measurement unit was constructed and validated. Subsequently, we investigated four different catheter models, each equipped with a unique measurement technology Tacticath Quartz (Abbott), AcQBlate Force (Biotronik/Acutus), Stablepoint (Boston Scientific), and Smarttouch SF (Biosense Webster). For each model, the accuracy of three different catheters was measured within the range of 0-60 g and at contact angles of 0°, 30°, 45°, 60°, and 90°.

RESULTS:

In total, 6685 measurements were performed using 4 × 3 catheters (median of 568, interquartile range 511-606 measurements per catheter). Over the entire measurement-range, the force measured by the catheters deviated from the real force by the following absolute mean values Tacticath 1.29 ± 0.99 g, AcQBlate Force 2.87 ± 2.37 g, Stablepoint 1.38 ± 1.29 g, and Smarttouch 2.26 ± 2.70 g. For some models, significant under- and overestimation of >10 g were observed at higher forces. Mean absolute errors of all models across the range of 10-40 g were <3 g.

CONCLUSION:

Contact measured by force-sensing catheters is accurate with 1-3 g deviation within the range of 10-40 g. Significant errors can occur at higher forces with potential clinical consequences.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Ablación por Catéter / Catéteres Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Suiza

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Ablación por Catéter / Catéteres Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Suiza