Your browser doesn't support javascript.
loading
Tubridge Flow Diverter for the Treatment of Unruptured Dissecting Cerebral Aneurysms.
Xie, Dajiang; Zhao, Li; Liu, Hailong; Wang, Jiaxiong; Lu, Peng; Ye, Xin; Yang, Shuxu.
Afiliación
  • Xie D; Department of Neurosurgery, Sir Run Run Shaw Hospital, College of Medical Sciences, Zhejiang University, Hangzhou, China.
  • Zhao L; Department of Gastroenterology, Hangzhou Red Cross Hospital/Integrated Traditional Chinese and Western Medicine Hospital of Zhejiang Province, Zhejiang Chinese Medical University, Hangzhou, China.
  • Liu H; Department of Neurosurgery, Jiashan Hospital of Traditional Chinese Medicine, Jiaxing, China.
  • Wang J; Department of Neurosurgery, South Yunnan Central Hospital of Yunnan Province (The First People's Hospital of HongHe Prefecture), Mengzi, China.
  • Lu P; Department of Neurosurgery, Sir Run Run Shaw Hospital, College of Medical Sciences, Zhejiang University, Hangzhou, China.
  • Ye X; Department of Neurosurgery, Sir Run Run Shaw Hospital, College of Medical Sciences, Zhejiang University, Hangzhou, China.
  • Yang S; Department of Neurosurgery, Sir Run Run Shaw Hospital, College of Medical Sciences, Zhejiang University, Hangzhou, China. Electronic address: siryangshuxu@163.com.
World Neurosurg ; 172: e343-e348, 2023 Apr.
Article en En | MEDLINE | ID: mdl-36639100
ABSTRACT

BACKGROUND:

The Tubridge flow diverter is a device widely used in China aimed at reconstructing parent artery and occluding complex aneurysm. The experience of the Tubridge in treating unruptured vertebrobasilar artery dissecting aneurysms is still limited. In this study, we aimed to evaluate the safety and efficacy of the Tubridge flow diverter for the treatment of vertebrobasilar artery dissecting aneurysms.

METHODS:

We reviewed the clinical records of aneurysms treated with the Tubridge flow diverter between 2019 and 2021 in a national cerebrovascular disease center. Therapeutic process, occlusion rate, and clinical outcome were compared.

RESULTS:

Twenty-three patients with 23 vertebrobasilar artery aneurysms were identified. The results showed that the mean length and mean maximal width were 15.14 and 9.14 mm, respectively, in the vertebrobasilar artery. Twenty-four Tubridge flow diverters were successfully implanted without unfold failure. A complete occlusion rate at the last angiographic follow-up was achieved in 78.26% of vertebrobasilar artery aneurysms. Fifteen branch arteries were covered, and only 1 branch artery disappeared at follow-up. Mild asymptomatic cerebral infarction occurred in 3 patients (13.04%); intracranial hemorrhage was not found in the patients.

CONCLUSIONS:

Our preliminary experience suggests that the Tubridge flow diverter might be a safe and effective tool for dissecting cerebral aneurysms. Branch arteries were well protected and mild asymptomatic cerebral infarction occurred in some patients. Adequate evidence is required to clear the definite indications and complications in a multicenter randomized controlled trial with a long-term follow-up.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Aneurisma Intracraneal / Embolización Terapéutica / Procedimientos Endovasculares / Disección Aórtica Tipo de estudio: Clinical_trials / Observational_studies Límite: Humans Idioma: En Revista: World Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2023 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Aneurisma Intracraneal / Embolización Terapéutica / Procedimientos Endovasculares / Disección Aórtica Tipo de estudio: Clinical_trials / Observational_studies Límite: Humans Idioma: En Revista: World Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2023 Tipo del documento: Article País de afiliación: China