Evaluation of a Fourth-Generation Subcutaneous Real-Time Continuous Glucose Monitor (CGM) in Individuals With Diabetes on Peritoneal Dialysis.
Ng, Jack K C; Ling, James; Luk, Andrea O Y; Lau, Eric S H; Ma, Ronald C W; Li, Philip K T; Szeto, Cheuk Chun; Chan, Juliana C N; Chow, Elaine.
Afiliación
  • Ng JKC; Department of Medicine and Therapeutics, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong SAR.
  • Ling J; Department of Medicine and Therapeutics, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong SAR.
  • Luk AOY; Department of Medicine and Therapeutics, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong SAR.
  • Lau ESH; Phase 1 Clinical Trial Centre, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong SAR.
  • Ma RCW; Hong Kong Institute of Diabetes and Obesity,Hong Kong SAR.
  • Li PKT; Department of Medicine and Therapeutics, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong SAR.
  • Szeto CC; Department of Medicine and Therapeutics, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong SAR.
  • Chan JCN; Hong Kong Institute of Diabetes and Obesity,Hong Kong SAR.
  • Chow E; Li Ka Shing Institute of Health Sciences, The Chinese University of Hong Kong, Hong Kong SAR.
Diabetes Care ; 46(6): 1191-1195, 2023 06 01.
Article en En | MEDLINE | ID: mdl-37043824
OBJECTIVE: To evaluate the performance of a real-time continuous glucose monitor (CGM) in individuals with diabetes on peritoneal dialysis (PD). RESEARCH DESIGN AND METHODS: Thirty participants with type 2 diabetes on continuous ambulatory PD wore a Guardian Sensor 3 on the upper arm paired with Guardian Connect for 14 days. We compared CGM readings against Yellow Springs Instrument (YSI) venous glucose during an 8-h in-clinic session with glucose challenge. RESULTS: The mean absolute relative difference (MARD) was 10.4% (95% CI 9.6, 11.7) from 941 CGM-YSI matched pairs; 81.3% of readings were within %15/15 of YSI values in the full glycemic range. Consensus error grid analysis showed 99.9% of sensor values in zones A and B. There were no correlations between pH, uremia, hydration status, and MARD. CONCLUSIONS: We showed satisfactory performance of a real-time CGM sensor in PD patients with diabetes, supporting future use to facilitate treatment decisions.