Radiation dose during interventional cardiology procedures: portable C-arm vs. a new generation fluoroscopy system.
Pierce, K V; Scansen, B A; Rao, S.
Afiliación
  • Pierce KV; Department of Clinical Sciences, Colorado State University, Fort Collins, CO 80523, USA. Electronic address: kvpierce@ncsu.edu.
  • Scansen BA; Department of Clinical Sciences, Colorado State University, Fort Collins, CO 80523, USA.
  • Rao S; Department of Clinical Sciences, Colorado State University, Fort Collins, CO 80523, USA.
J Vet Cardiol ; 47: 30-40, 2023 Jun.
Article en En | MEDLINE | ID: mdl-37150018
ABSTRACT

INTRODUCTION:

Occupational exposure to ionizing radiation poses health risks for veterinary interventionalists. There are limited veterinary studies evaluating radiation dose in the cardiac catheterization laboratory. The purpose of this study was to report direct radiation dose exposure to patients during common interventional cardiology procedures and compare these doses between two fluoroscopy units. ANIMALS One hundred and fifty-four client-owned dogs. MATERIALS AND

METHODS:

Patient dose during procedures using a portable C-arm were retrospectively analyzed and compared to those performed in a contemporary interventional suite. Fluoroscopy equipment, procedure type, operator, patient weight, fluoroscopy time, dose area product, and air kerma were recorded and statistically modeled using univariable and multivariable linear regression to evaluate the effect of each factor.

RESULTS:

Patient dose population (154 dogs), comprised 61 patent ductus arteriosus occlusions, 60 balloon pulmonary valvuloplasties, and 33 pacemaker implantations. Patient dose was significantly lower in the group utilizing a newer generation fluoroscopy unit vs. the group utilizing an older portable C-arm, positively correlated with patient weight, and highest during balloon pulmonary valvuloplasties compared to patent ductus arteriosus occlusions or pacemaker implantations (all P<0.010).

DISCUSSION:

Newer fluoroscopy systems can be equipped with technologies that improve image quality while reducing patient dose and radiation exposure to interventional personnel.

CONCLUSIONS:

We documented a significant reduction in patient radiation dose using a newer fluoroscopy system as compared to an older portable C-arm for interventional cardiology procedures in animals. Improved knowledge of patient radiation dose factors may promote better radiation safety protocols in veterinary interventional cardiology.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Enfermedades de los Perros / Conducto Arterioso Permeable Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Animals Idioma: En Revista: J Vet Cardiol Asunto de la revista: CARDIOLOGIA / MEDICINA VETERINARIA Año: 2023 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Enfermedades de los Perros / Conducto Arterioso Permeable Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Animals Idioma: En Revista: J Vet Cardiol Asunto de la revista: CARDIOLOGIA / MEDICINA VETERINARIA Año: 2023 Tipo del documento: Article