An 8 mm endoscopic histotripsy array with integrated high-resolution ultrasound imaging.
Mallay, Matthew G; Landry, Thomas G; Brown, Jeremy A.
Afiliación
  • Mallay MG; School of Biomedical Engineering, Dalhousie University, Halifax, NS, Canada. Electronic address: matt.mallay@dal.ca.
  • Landry TG; School of Biomedical Engineering, Dalhousie University, Halifax, NS, Canada.
  • Brown JA; School of Biomedical Engineering, Dalhousie University, Halifax, NS, Canada; Department of Electrical and Computer Engineering, Dalhousie University, Halifax, NS, Canada.
Ultrasonics ; 139: 107275, 2024 Apr.
Article en En | MEDLINE | ID: mdl-38508082
ABSTRACT
An 8 mm diameter, image-guided, annular array histotripsy transducer was fabricated and characterized. The array was laser etched on a 5 MHz, 1-3 dice and fill, PZT-5H/epoxy composite with a 45 % volume fraction. Flexible PCBs were used to electrically connect to the array elements using wirebonds. The array was backed with a low acoustic impedance epoxy mixture. A 3.6 by 3.8 mm, 64-element, 30 MHz phased array imaging probe was positioned in the center hole, to co-align the imaging plane with the bubble cloud produced by the therapy array. A custom 16-channel high voltage pulse generator was used to test the annular array for focal lengths ranging from 3- to 8-mm. An aluminum lens-focussed transducer with a 7 mm focal length was fabricated using the same piezocomposite and backing material and tested along with the histotripsy array. Simulated results from COMSOL FEM models were compared to measured results for low voltage characterization of the array and lens-focussed transducer. The measured transmit sensitivity of the array ranged from 0.113 to 0.167 MPa/V, while the lens-focussed transducer was 0.192 MPa/V. Simulated values were 0.160 to 0.174 MPa/V and 0.169 MPa/V, respectively. The measured acoustic fields showed a significantly increased depth-of-field compared the lens-focussed transducer, while the beamwidths of the array focus were comparable to the lens. The measured cavitation voltage in water was between 254 V and 498 V depending on the focal length, and 336 V for the lens-focussed transducer. The array had a lower cavitation voltage than the lens-focussed transducer for a comparable operating depth. The histotripsy array was tested in a tissue phantom and an in vivo rat brain. It was used to produce an elongated lesion in the brain by electronically steering the focal length from 3- to 8-mm axially. Real time ultrasound imaging with a Doppler overlay was used to target the tissue and monitor ablation progress, and histology confirmed the targeted tissue was fully homogenized.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Ultrasonido Enfocado de Alta Intensidad de Ablación Idioma: En Revista: Ultrasonics Año: 2024 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Ultrasonido Enfocado de Alta Intensidad de Ablación Idioma: En Revista: Ultrasonics Año: 2024 Tipo del documento: Article