Feasibility Analysis of Individualized Low Flow Rate Abdominal Contrast-Enhanced Computed Tomography in Chemotherapy Patients: Dual-Source Computed Tomography With Low Tube Voltage.
Zhang, Yicun; Yuan, Dian; Qi, Ke; Zhang, Mengyuan; Zhang, Weiting; Wei, Nannan; Li, Linfeng; Lv, Peijie; Gao, Jianbo; Liu, Jie.
Afiliación
  • Zhang Y; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Yuan D; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Qi K; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Zhang M; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Zhang W; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Wei N; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Li L; Siemens Healthineers, Shanghai, China.
  • Lv P; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Gao J; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
  • Liu J; From the The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou.
Article en En | MEDLINE | ID: mdl-38693081
ABSTRACT

PURPOSE:

The aim of the study is to investigate the feasibility of using dual-source computed tomography (CT) combined with low flow rate and low tube voltage for postchemotherapy image assessment in cancer patients.

METHODS:

Ninety patients undergoing contrast-enhanced CT scans of the upper abdomen were prospectively enrolled and randomly assigned to groups A, B, and C (n = 30 each). In group A, patients underwent scans at 120 kVp with 448 mgI/kg. Patients in group B underwent scans at 100 kVp with 336 mgI/kg. Patient in group C underwent scans at 70 kVp with of 224 mgI/kg. Quantitative measurements including the CT number, standard deviation of CT number, signal-to-noise ratio, contrast-to-noise ratio, subjective reader scores, and the volume and flow rate of contrast agent were evaluated for each group.

RESULTS:

There was no statistically significant difference in the subjective image scores within the three groups except for the kidney (all P > 0.05). Group C showed significantly higher CT values, lower noise levels, and higher signal-to-noise ratio and contrast-to-noise ratio values in the majority of the regions of interest compared to the other groups (P < 0.05). In group C, the contrast agent dose was decreased by 46% compared to group A (79.48 ± 12.24 vs 42.7 ± 8.6, P < 0.01), and the contrast agent injection rate was reduced by 22% (2.7 ± 0.41 vs 2.1 ± 0.4, P < 0.01).

CONCLUSIONS:

The use of 70 kVp tube voltage combined with low iodine flow rates prove to be a more effective approach in solving the challenge of compromised blood vessels in postchemotherapy tumor patients, without reducing image quality and diagnostic confidence.