Crystallization on intraocular lens surfaces associated with the use of Healon GV.
Arch Ophthalmol
; 112(8): 1037-42, 1994 Aug.
Article
en En
| MEDLINE
| ID: mdl-8053816
OBJECTIVE: To describe a series of patients who had visually significant crystalline deposits on their intraocular lenses during cataract surgery with the use of Healon GV (a high concentration and high molecular-weight hyaluronate sodium). METHODS: Patients were examined for crystalline deposits on their intraocular lenses. These deposits were compared with intraocular lens type, viscoelastic solutions, any other intraocular substance used, type of surgery, and perioperative medications. RESULTS: In the 11 patients with documented changes (six photographically), the only consistent finding was the use of Healon GV. Furthermore, since we discontinued the use of Healon GV, we have not seen a recurrence of these deposits in more than 500 consecutive patients. The deposits could last a long time (at least 6 months) if sequestered by the posterior capsule, and they are believed to be visually significant at times (Snellen visual acuity of 20/40 or worse). CONCLUSION: Healon GV use is associated with a new clinical finding of crystalline deposits on intraocular lenses. These deposits can be clinically significant.
Buscar en Google
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Complicaciones Posoperatorias
/
Ácido Hialurónico
/
Lentes Intraoculares
Tipo de estudio:
Risk_factors_studies
Límite:
Aged
/
Female
/
Humans
/
Male
Idioma:
En
Revista:
Arch Ophthalmol
Año:
1994
Tipo del documento:
Article