Your browser doesn't support javascript.
loading
Analysis and Processing of Flow Alarm about Fresenius 4008 S Hemodialysis Machine / 中国医疗器械杂志
Article en Zh | WPRIM | ID: wpr-942744
Biblioteca responsable: WPRO
ABSTRACT
The Fresenius 4008 S hemodialysis machine is one of the main hemodialysis machines in our hospital. Because of its longer service life, the probability of malfunction is larger. For flow alarm is the typical fault of this model, we combined with practical maintenance experience and theoretical technology, then briefly analyzed and summarized the malfunction. On the principle of easy to difficult, we check in order from five aspects, they are pipeline blockage, aging of the degassing pump or flow pump, the float switch malfunction, valve abnormality and the control circuit problem. Through this one-by-one and systematic troubleshooting, the flow alarm can usually be solved, also we can save the maintenance time and the cost.
Asunto(s)
Palabras clave
Texto completo: 1 Banco de datos: WPRIM Asunto principal: Diálisis Renal / Mantenimiento Idioma: Zh Revista: Chinese Journal of Medical Instrumentation Año: 2020 Tipo del documento: Article
Texto completo: 1 Banco de datos: WPRIM Asunto principal: Diálisis Renal / Mantenimiento Idioma: Zh Revista: Chinese Journal of Medical Instrumentation Año: 2020 Tipo del documento: Article