1.
J Hosp Infect ; 80(3): 262-4, 2012 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-22277192

RESUMEN

We developed a percutaneous electrical transducer for home therapy of chronic pain, a device that requires high level disinfection between uses. The utility of boiling water to provide high level disinfection was evaluated by inoculating transducer pads with potential skin pathogens (Staphylococcus aureus, Mycobacterium terrae, Pseudomonas aeruginosa, Candida albicans) and subjecting them to full immersion in water boiling at 4200 feet elevation (95 °C). Log10 reductions in colony-forming units (cfu) at 10 min were 7.1, >6.3 and >5.5 for S. aureus, P. aeruginosa and C. albicans, respectively, but only 4.6 for M. terrae. At 15 min the reductions had increased to 7.5, >6.8, >6.6 and >7.5 cfu, respectively.


Asunto(s)
Dolor Crónico/terapia , Desinfección/métodos , Equipos y Suministros , Servicios de Atención de Salud a Domicilio , Calor , Enfermedades Cutáneas Bacterianas/prevención & control , Transductores/microbiología , Agua , Candida albicans/crecimiento & desarrollo , Candidiasis/microbiología , Candidiasis/prevención & control , Humanos , Micobacterias no Tuberculosas/crecimiento & desarrollo , Pseudomonas aeruginosa/crecimiento & desarrollo , Piel/microbiología , Enfermedades Cutáneas Bacterianas/microbiología , Staphylococcus aureus/crecimiento & desarrollo
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA