Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
Más filtros

Base de datos
Tipo de estudio
Tipo del documento
Intervalo de año de publicación
1.
Rev Sci Instrum ; 92(5): 053529, 2021 May 01.
Artículo en Inglés | MEDLINE | ID: mdl-34243238

RESUMEN

The Joint European Torus (JET) gamma-ray camera has been recently upgraded with the installation of new gamma-ray detectors, based on LaBr3(Ce) scintillation crystals, which add spectroscopic capability to the existing system allowing measurements with good energy resolution (5% at 0.622 MeV), a dynamic range from hundreds of keV up to about 30 MeV, and high counting rate capabilities of MCps. First gamma-ray measurements during the C38 campaign of the JET have been successfully carried out, in particular, in D-3He plasmas from three-ion ion cyclotron resonance heating experiments, where the detection of 16.4 MeV γ-rays from D + 3He → γ + 5Li reactions with the gamma-ray camera upgrade allowed determining the spatial profile of alpha particles born in D + 3He fusion reactions.

2.
Rev Sci Instrum ; 89(10): 10I116, 2018 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-30399923

RESUMEN

The JET gamma-ray cameras have been recently upgraded within the gamma-ray camera upgrade project in support of development of JET high performance deuterium plasma scenarios and in preparation of deuterium-tritium experiments. New, dedicated detectors based on a LaBr3 crystal and silicon photo-multipliers have been developed and replaced pre-existing CsI detectors in all 19 channels. The new instrument gives opportunity of making two-dimensional gamma-ray measurements with a counting rate capability exceeding 1 MCounts/s (MCps) and energy resolution better than 5% at 1.1 MeV. The upgrade is of relevance for fast ion and runaway electron physics studies in high performance deuterium discharges and also in plasmas with tritium at neutron yields in the range up to about 5 × 1017 n/s.

3.
Rev Sci Instrum ; 85(4): 043506, 2014 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-24784606

RESUMEN

First simultaneous measurements of deuterium-deuterium (DD) and deuterium-tritium neutrons from deuterium plasmas using a Single crystal Diamond Detector are presented in this paper. The measurements were performed at JET with a dedicated electronic chain that combined high count rate capabilities and high energy resolution. The deposited energy spectrum from DD neutrons was successfully reproduced by means of Monte Carlo calculations of the detector response function and simulations of neutron emission from the plasma, including background contributions. The reported results are of relevance for the development of compact neutron detectors with spectroscopy capabilities for installation in camera systems of present and future high power fusion experiments.

4.
Rev Sci Instrum ; 83(10): 10D505, 2012 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-23130789

RESUMEN

An in-vessel calibration light source (ICLS) has been implemented for remote use during extended shutdown periods of the Joint European Torus (JET). The ICLS facilitated the in situ calibration of optical diagnostics, which previously were performed when the diagnostics were removed from JET. Since the ICLS is used to calibrate diagnostics over the entire, exact optical path as used when plasma discharge data are measured, the ICLS calibration implicitly accounts for any vignetting losses in the JET vessel viewports in addition to the vacuum window transmission. At least ten diagnostic systems have benefited from the ICLS during the extended ITER-like wall shutdown of 2009-2011. Examples of the use of the ICLS in JET are given.


Asunto(s)
Luz , Fenómenos Magnéticos , Física/instrumentación , Calibración , Gases em Plasma/química
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA