Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros

Base de datos
Tipo del documento
País de afiliación
Intervalo de año de publicación
1.
Strahlenther Onkol ; 2024 Jul 05.
Artículo en Inglés | MEDLINE | ID: mdl-38967820

RESUMEN

PURPOSE: A prototype infrared camera - cone-beam computed tomography (CBCT) system for tracking in brachytherapy has recently been developed. We evaluated for the first time the corresponding tracking accuracy and uncertainties, and implemented a tracking-based prediction of needles on CBCT scans. METHODS: A marker tool rigidly attached to needles was 3D printed. The precision and accuracy of tool tracking was then evaluated for both static and dynamic scenarios. Euclidean distances between the tracked and CBCT-derived markers were assessed as well. To implement needle tracking, ground truth models of the tool attached to 200 mm and 160 mm needles were matched to the tracked positions in order to project the needles into CBCT scans. Deviations between projected and actual needle tips were measured. Finally, we put our results into perspective with simulations of the system's tracking uncertainties. RESULTS: For the stationary scenario and dynamic movements, we achieved tool-tracking precision and accuracy of 0.04 ± 0.06 mm and 0.16 ± 0.18 mm, respectively. The tracked marker positions differed by 0.52 ± 0.18 mm from the positions determined via CBCT. In addition, the predicted needle tips in air deviated from the actual tip positions by only 1.62 ± 0.68 mm (200 mm needle) and 1.49 ± 0.62 mm (160 mm needle). The simulated tracking uncertainties resulted in tip variations of 1.58 ± 0.91 mm and 1.31 ± 0.69 mm for the 200 mm and 160 mm needles, respectively. CONCLUSION: With the innovative system it was possible to achieve a high tracking and prediction accuracy of marker tool and needles. The system shows high potential for applicator tracking in brachytherapy.

2.
J Appl Clin Med Phys ; 25(7): e14364, 2024 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-38626753

RESUMEN

PURPOSE: To enable a real-time applicator guidance for brachytherapy, we used for the first time infra-red tracking cameras (OptiTrack, USA) integrated into a mobile cone-beam computed tomography (CBCT) scanner (medPhoton, Austria). We provide the first description of this prototype and its performance evaluation. METHODS: We performed assessments of camera calibration and camera-CBCT registration using a geometric calibration phantom. For this purpose, we first evaluated the effects of intrinsic parameters such as camera temperature or gantry rotations on the tracked marker positions. Afterward, calibrations with various settings (sample number, field of view coverage, calibration directions, calibration distances, and lighting conditions) were performed to identify the requirements for achieving maximum tracking accuracy based on an in-house phantom. The corresponding effects on camera-CBCT registration were determined as well by comparing tracked marker positions to the positions determined via CBCT. Long-term stability was assessed by comparing tracking and a ground-truth on a weekly basis for 6 weeks. RESULTS: Robust tracking with positional drifts of 0.02 ± 0.01 mm was feasible using the system after a warm-up period of 90 min. However, gantry rotations affected the tracking and led to inaccuracies of up to 0.70 mm. We identified that 4000 samples and full coverage were required to ensure a robust determination of marker positions and camera-CBCT registration with geometric deviations of 0.18 ± 0.03 mm and 0.42 ± 0.07 mm, respectively. Long-term stability showed deviations of more than two standard deviations from the initial calibration after 3 weeks. CONCLUSION: We implemented for the first time a standalone combined camera-CBCT system for tracking in brachytherapy. The system showed high potential for establishing corresponding workflows.


Asunto(s)
Braquiterapia , Tomografía Computarizada de Haz Cónico , Fantasmas de Imagen , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador , Radioterapia Guiada por Imagen , Humanos , Tomografía Computarizada de Haz Cónico/métodos , Tomografía Computarizada de Haz Cónico/instrumentación , Braquiterapia/instrumentación , Braquiterapia/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia Guiada por Imagen/métodos , Radioterapia Guiada por Imagen/instrumentación , Calibración , Procesamiento de Imagen Asistido por Computador/métodos , Radioterapia de Intensidad Modulada/métodos , Neoplasias/radioterapia , Neoplasias/diagnóstico por imagen
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA