Clin Biochem ; 41(1-2): 10-8, 2008 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-17988658

RESUMEN

OBJECTIVES: To evaluate the inter-laboratory reproducibility of blood test for liver fibrosis: FibroMeter, Fibrotest, APRI and their composites variables. DESIGN AND METHODS: Four studies, including 147 patients, were performed: study #1 included 2 metachronous blood samples and 2 laboratories; studies #2, #3 and #4 included synchronous samples with assays delayed at day 1 in 12 laboratories, at day 0 in 10 laboratories and at day 0 or 1 in 2 laboratories, respectively. Agreement was evaluated by the intraclass correlation coefficient (r(ic)). RESULTS: In studies #1, #2 and #4, r(ic) for FibroMeter was 0.893, 0.942 and 0.991, respectively. In study #3, the r(ic) were: FibroMeter: 0.963, Fibrotest: 0.984, APRI: 0.949. Large simulated variations in composite variables had a weak impact on FibroMeter. CONCLUSIONS: When blood marker limits are controlled, inter-laboratory agreement of blood tests is excellent in clinical practice conditions. Blood tests are robust against the variability of composite blood variables.


Asunto(s)
Cirrosis Hepática/sangre , Pruebas de Función Hepática , Adulto , Anciano , Femenino , Humanos , Cirrosis Hepática/diagnóstico , Cirrosis Hepática/patología , Masculino , Persona de Mediana Edad , Variaciones Dependientes del Observador , Reproducibilidad de los Resultados