Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros

Base de datos
Tipo del documento
País de afiliación
Intervalo de año de publicación
1.
Micromachines (Basel) ; 15(2)2024 Jan 31.
Artículo en Inglés | MEDLINE | ID: mdl-38398953

RESUMEN

In this paper, a fiber optic microprobe displacement sensor is proposed considering characteristics of micro-Michelson interference structure and its components. The principal error of micro Fabry-Perot interferometric structure is avoided, and high-precision interferometric displacement measurement is realized. The collimated microprobe and convergent microprobe are analyzed, simulated, and designed for the purposes of measuring long-distance displacement and small spot rough surface, respectively. The core parameters of the probes' internal components are mapped to coupling efficiency and contrast of the sensor measurements, which provides a basis for the probes' design. Finally, simulation and experimental testing of the two probes show that the collimated probe's working distance and converging probe's tolerance angle can reach 40 cm and ±0.5°, respectively. The designed probes are installed in the fiber laser interferometer, and a displacement resolution of 0.4 nm is achieved.