An automated and modified technique for testing the retinal function (Arden test) by use of the electro-oculogram (EOG) for clinical and research use.
Behrens, F; Weiss, L R.
Afiliación
  • Behrens F; Universitätsklinikum Benjamin Franklin, Freie Universität Berlin, Institut für Physiologie, Germany. Behrens2@zedat.fu-berlin.de
Doc Ophthalmol ; 96(4): 283-92, 1998.
Article en En | MEDLINE | ID: mdl-10855804
ABSTRACT
The electro-oculogram (EOG) can be recorded by use of skin electrodes positioned near the eye in the horizontal or vertical plane and is dependent on the position of the eye relative to the electrodes and the size of the corneo-retinal potential (CRP). By use of the functional relation between the recorded potential and the eye position - CRP is considered to be constant - the EOG is the eye movement recording method most often used. Changes in CRP dependent on luminance or metabolic influences can be measured, on the other hand, by means of the EOG when eye movements of constant amplitudes are performed. The EOG in this paper is used in the latter sense. The widely used clinical test of rentinal function based on the electro-oculogram firstly described by Arden (Arden index) has been modified to achieve more precise and reliable results by including a gradual dark adaptation procedure and a special data analysis to detect fixation periods. Using modern techniques, the complete procedure has been automated. The efficiency of the procedure and the device is demonstrated in various ways.
Asunto(s)
Buscar en Google
Base de datos: MEDLINE Asunto principal: Retina / Electrooculografía Límite: Humans Idioma: En Revista: Doc Ophthalmol Año: 1998 Tipo del documento: Article País de afiliación: Alemania
Buscar en Google
Base de datos: MEDLINE Asunto principal: Retina / Electrooculografía Límite: Humans Idioma: En Revista: Doc Ophthalmol Año: 1998 Tipo del documento: Article País de afiliación: Alemania