Absorbed radiation dose of the female breast during diagnostic multidetector chest CT and dose reduction with a tungsten-antimony composite breast shield: preliminary results.
Parker, M S; Kelleher, N M; Hoots, J A; Chung, J K; Fatouros, P P; Benedict, S H.
Afiliación
  • Parker MS; Medical College of Virginia Hospitals-VCU Health System, Richmond, VA 23298-0615, USA. msparker@mail1.vcu.edu
Clin Radiol ; 63(3): 278-88, 2008 Mar.
Article en En | MEDLINE | ID: mdl-18275868
ABSTRACT

AIM:

To determine the absorbed radiation dose to the female breast during chest computed tomography (CT), and whether a custom-designed breast shield can reduce that dose. MATERIALS AND

METHODS:

Bilateral breast phantoms were combined with an anthropomorphic torso phantom. Each breast phantom contained 20 thermoluminescent dosimeter (TLD) cavities. Eight cavities were used per phantom. Absorbed radiation was measured using TLD 100 s. Three-stacked TLDs comprised a set. Three sets of three TLDs were positioned at eight designated locations and three depths (surface; 1 cm; 4 cm). One set of three TLDs was positioned at eight additional designations, 1cm deep. Each breast was divided anatomically into quadrants. In total, 32 TLD sets/96 TLDs were deployed. The breast-torso phantom was consecutively imaged using a 16-detector array CT machine. Subsequently, 32 new TLD sets were similarly placed, the phantom re-imaged in a likewise manner, but with the application of a tungsten-antimony composite breast shield. TLD readings were averaged and calculated.

RESULTS:

Average absorbed radiation doses for unshielded right and left breast phantoms ranged from 13.83-19.36 mGy, and 14-20.47 mGy, respectively. The absorbed dose in the shielded right and left breast was reduced to 6.64-8.12 mGy, and 6.7-8.03 mGy, respectively. Average absorbed radiation doses based on the depth for the unshielded breasts ranged from 15.4-18.3 mGy. Shielding reduced this dose to 7-7.9 mGy. Unshielded absorbed radiation doses based on anatomic quadrants ranged from 17.5-18.9 mGy. Shielding reduced this dose to 7-7.5 mGy.

CONCLUSIONS:

The average absorbed radiation dose to the unshielded female breast phantom is approximately 14-20 mGy. An externally applied shield can reduce this absorbed dose by 56-61%.
Asunto(s)

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Equipos de Seguridad / Protección Radiológica / Mama / Mamografía / Tomografía Computarizada por Rayos X Tipo de estudio: Diagnostic_studies Límite: Female / Humans Idioma: En Revista: Clin Radiol Año: 2008 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Equipos de Seguridad / Protección Radiológica / Mama / Mamografía / Tomografía Computarizada por Rayos X Tipo de estudio: Diagnostic_studies Límite: Female / Humans Idioma: En Revista: Clin Radiol Año: 2008 Tipo del documento: Article País de afiliación: Estados Unidos