Your browser doesn't support javascript.
loading
Hydroxyapatite-coated pins versus titanium alloy pins in external fixation at the wrist: a controlled cohort study.
Pieske, Oliver; Pichlmaier, Leopold; Kaltenhauser, Felicitas; Schramm, Nicolai; Rubenbauer, Bianka; Greiner, Axel; Piltz, Stefan.
Afiliación
  • Pieske O; Department of Trauma Surgery, University Hospital of Munich, Marchioninistraße, Munich, Germany. Oliver.Pieske@med.uni-muenchen.de
J Trauma ; 70(4): 845-51, 2011 Apr.
Article en En | MEDLINE | ID: mdl-20805762
ABSTRACT

BACKGROUND:

The purpose of this study was to analyze whether the prevalence of pin-related complications can be reduced by the use of hydroxyapatite (HA)-coated pins in external fixators applied for unstable wrist fractures.

METHODS:

Forty patients (160 pins) were randomized for standard uniplanar fixator treatment with the use of identically designed pins either composed of titanium-alloy (Ti6Al4V) (n = 20) or coated by HA (n = 20). Each pin site was clinically evaluated with regard to erythema, drainage, pain value, and radiologically assessed concerning loosening at T1 (mean, 9 days), T2 (mean, 43 days), and T3 (mean, 56 days). In case of pin-track complication, the patient was followed continuously. The need for antibiotics or additional surgery was documented. Bone mineral density was analyzed by Dual Energy X-ray Absorptiometry. At fixator removal (T2), the pin-extraction strength was measured by the use of a digital-torque-wrench.

RESULTS:

Two minor pin-track infections requiring oral antibiotics occurred in the HA-pin group (2.7%) (p > 0.05). The vast majority of clinical pin-site parameters were comparable in both groups. At the end of the fixator therapy, there were 16 loose pins (n(Ti6AL4V-group) = 10; n(HA-group) = 6). The rate of loose pins was correlated to patient's age (p < 0.05) but not to bone mineral density values or the occurrence of pin-site infection. Finally, no significant difference between the two groups was detected with regard to the prevalence of clinical relevant pin-site complications (p = 0.80).

CONCLUSIONS:

In external fixation of the wrist, the use of HA-coated pins yields no clinical advantages there is a trend toward a superior pin-bone anchorage, but a tendency of increased susceptibility for minor pin-track infections.
Asunto(s)

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Traumatismos de la Muñeca / Clavos Ortopédicos / Durapatita / Materiales Biocompatibles Revestidos / Aleaciones / Fracturas Óseas / Fijación de Fractura Tipo de estudio: Clinical_trials / Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Trauma Año: 2011 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Traumatismos de la Muñeca / Clavos Ortopédicos / Durapatita / Materiales Biocompatibles Revestidos / Aleaciones / Fracturas Óseas / Fijación de Fractura Tipo de estudio: Clinical_trials / Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Trauma Año: 2011 Tipo del documento: Article País de afiliación: Alemania