A simulator for oscillometric blood-pressure signals to test automated noninvasive sphygmomanometers.
Riedel, W; Mieke, S; Seemann, R; Ittermann, B.
Afiliación
  • Riedel W; Physikalisch-Technische Bundesanstalt, Berlin, Germany.
Rev Sci Instrum ; 82(2): 024303, 2011 Feb.
Article en En | MEDLINE | ID: mdl-21361618
A device was developed allowing to generate simulated human blood pressure signals for the purpose of testing the performance of automated noninvasive sphygmomanometers. The apparatus reproducibly generates blood-pressure oscillations synthesized from prerecorded measurements on human subjects. These real-life data allow for a much better evaluation of the accuracy of blood-pressure measurements than the existing simulators using artificial and thus less realistic waveforms. To assess the performance of a given sphygmomanometer under both stable and varying conditions, generated signals can be repeated in their original shape or distorted by well-defined artifacts. In comparison to clinical tests, the procedural influences on the performance testing of sphygmomanometers are largely reduced when the simulator is used.
Asunto(s)

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Oscilometría / Presión Sanguínea / Procesamiento de Señales Asistido por Computador / Esfigmomanometros Límite: Humans Idioma: En Revista: Rev Sci Instrum Año: 2011 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Oscilometría / Presión Sanguínea / Procesamiento de Señales Asistido por Computador / Esfigmomanometros Límite: Humans Idioma: En Revista: Rev Sci Instrum Año: 2011 Tipo del documento: Article País de afiliación: Alemania