Highly sensitive covalently functionalised integrated silicon nanowire biosensor devices for detection of cancer risk biomarker.
Mohd Azmi, M A; Tehrani, Z; Lewis, R P; Walker, K-A D; Jones, D R; Daniels, D R; Doak, S H; Guy, O J.
Afiliación
  • Mohd Azmi MA; College of Engineering, Swansea University, Singleton Park, Swansea, UK; University Kuala Lumpur, British Malaysian Institute, Gombak, Selangor, Malaysia. Electronic address: M.A.MOHD-AZMI.400194@swansea.ac.uk.
Biosens Bioelectron ; 52: 216-24, 2014 Feb 15.
Article en En | MEDLINE | ID: mdl-24060972
In this article we present ultra-sensitive, silicon nanowire (SiNW)-based biosensor devices for the detection of disease biomarkers. An electrochemically induced functionalisation method has been employed to graft antibodies targeted against the prostate cancer risk biomarker 8-hydroxydeoxyguanosine (8-OHdG) to SiNW surfaces. The antibody-functionalised SiNW sensor has been used to detect binding of the 8-OHdG biomarker to the SiNW surface within seconds of exposure. Detection of 8-OHdG concentrations as low as 1 ng/ml (3.5 nM) has been demonstrated. The active device has been bonded to a disposable printed circuit which can be inserted into an electronic readout system as part of an integrated Point of Care (POC) diagnostic. The speed, sensitivity and ease of detection of biomarkers using SiNW sensors render them ideal for eventual POC diagnostics.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Neoplasias de la Próstata / Técnicas Biosensibles / Desoxiguanosina / Nanocables Tipo de estudio: Diagnostic_studies / Etiology_studies / Risk_factors_studies Límite: Humans / Male Idioma: En Revista: Biosens Bioelectron Asunto de la revista: BIOTECNOLOGIA Año: 2014 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Neoplasias de la Próstata / Técnicas Biosensibles / Desoxiguanosina / Nanocables Tipo de estudio: Diagnostic_studies / Etiology_studies / Risk_factors_studies Límite: Humans / Male Idioma: En Revista: Biosens Bioelectron Asunto de la revista: BIOTECNOLOGIA Año: 2014 Tipo del documento: Article