Highly sensitive covalently functionalised integrated silicon nanowire biosensor devices for detection of cancer risk biomarker.
Biosens Bioelectron
; 52: 216-24, 2014 Feb 15.
Article
en En
| MEDLINE
| ID: mdl-24060972
In this article we present ultra-sensitive, silicon nanowire (SiNW)-based biosensor devices for the detection of disease biomarkers. An electrochemically induced functionalisation method has been employed to graft antibodies targeted against the prostate cancer risk biomarker 8-hydroxydeoxyguanosine (8-OHdG) to SiNW surfaces. The antibody-functionalised SiNW sensor has been used to detect binding of the 8-OHdG biomarker to the SiNW surface within seconds of exposure. Detection of 8-OHdG concentrations as low as 1 ng/ml (3.5 nM) has been demonstrated. The active device has been bonded to a disposable printed circuit which can be inserted into an electronic readout system as part of an integrated Point of Care (POC) diagnostic. The speed, sensitivity and ease of detection of biomarkers using SiNW sensors render them ideal for eventual POC diagnostics.
Palabras clave
Texto completo:
1
Base de datos:
MEDLINE
Asunto principal:
Neoplasias de la Próstata
/
Técnicas Biosensibles
/
Desoxiguanosina
/
Nanocables
Tipo de estudio:
Diagnostic_studies
/
Etiology_studies
/
Risk_factors_studies
Límite:
Humans
/
Male
Idioma:
En
Revista:
Biosens Bioelectron
Asunto de la revista:
BIOTECNOLOGIA
Año:
2014
Tipo del documento:
Article