Your browser doesn't support javascript.
loading
Alpha-defensin and the Synovasure lateral flow device for the diagnosis of prosthetic joint infection: a systematic review and meta-analysis.
Marson, B A; Deshmukh, S R; Grindlay, D J C; Scammell, B E.
Afiliación
  • Marson BA; Division of Rheumatology, Orthopaedics and Dermatology, University of Nottingham, Queen's Medical Centre, Nottingham, UK.
  • Deshmukh SR; Nottingham Elective Orthopaedic Service, Nottingham University Hospitals NHS Trust, City Hospital Campus, Nottingham, UK.
  • Grindlay DJC; Division of Rheumatology, Orthopaedics and Dermatology, University of Nottingham, Queen's Medical Centre, Nottingham, UK.
  • Scammell BE; Division of Rheumatology, Orthopaedics and Dermatology, University of Nottingham, Queen's Medical Centre, Nottingham, UK.
Bone Joint J ; 100-B(6): 703-711, 2018 06 01.
Article en En | MEDLINE | ID: mdl-29855233
ABSTRACT

Aims:

The aim of this review was to evaluate the available literature and to calculate the pooled sensitivity and specificity for the different alpha-defensin test systems that may be used to diagnose prosthetic joint infection (PJI). Materials and

Methods:

Studies using alpha-defensin or Synovasure (Zimmer Biomet, Warsaw, Indiana) to diagnose PJI were identified from systematic searches of electronic databases. The quality of the studies was evaluated using the Quality Assessment of Studies of Diagnostic Accuracy (QUADAS) tool. Meta-analysis was completed using a bivariate model.

Results:

A total of 11 eligible studies were included. The median QUADAS score was 13 (interquartile range 13 to 13) out of 14. Significant conflicts of interest were identified in five studies. The pooled sensitivity for the laboratory alpha-defensin test was 0.95 (95% confidence interval (CI) 0.91 to 0.98) and the pooled specificity was 0.97 (95% CI 0.95 to 0.98) for four studies with a threshold level of 5.2 mgl-1 The pooled sensitivity for the lateral flow cassette test was 0.85 (95% CI 0.74 to 0.92) and the pooled specificity was 0.90 (95% CI 0.91 to 0.98). There was a statistically significant difference in sensitivity (p = 0.019), but not specificity (p = 0.47).

Conclusion:

Laboratory-based alpha-defensin testing remains a promising tool for diagnosing PJI. The lateral flow cassette has a significantly lower performance and pooled results are comparable to the leucocyte esterase test. Further studies are required before the widespread adoption of the lateral flow cassette alpha-defensin test. Cite this article Bone Joint J 2018;100-B703-11.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Artroplastia / Biomarcadores / Infecciones Relacionadas con Prótesis / Alfa-Defensinas Tipo de estudio: Diagnostic_studies / Prognostic_studies / Systematic_reviews Límite: Humans Idioma: En Revista: Bone Joint J Año: 2018 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Artroplastia / Biomarcadores / Infecciones Relacionadas con Prótesis / Alfa-Defensinas Tipo de estudio: Diagnostic_studies / Prognostic_studies / Systematic_reviews Límite: Humans Idioma: En Revista: Bone Joint J Año: 2018 Tipo del documento: Article País de afiliación: Reino Unido