Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL.
Kuwabara, Mitsuo; Harada, Kanako; Hishiki, Yukiko; Kario, Kazuomi.
Afiliación
  • Kuwabara M; Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine, Tochigi, Japan.
  • Harada K; Omron Healthcare Co., Ltd., Kyoto, Japan.
  • Hishiki Y; Omron Healthcare Co., Ltd., Kyoto, Japan.
  • Kario K; Omron Healthcare Co., Ltd., Kyoto, Japan.
J Clin Hypertens (Greenwich) ; 21(6): 853-858, 2019 06.
Article en En | MEDLINE | ID: mdl-30803128
ABSTRACT
There is growing evidence of the clinical significance of daytime masked hypertension (MHT) and blood pressure (BP) variability (BPV). Recently, watch-type wearable devices for self-BP measurement have become available. Such devices might be promising tools to identify patients with daytime MHT or large BPV in their real-life conditions. The present study aimed to validate the accuracy of the Omron HEM-6410T-ZM and the Omron HEM-6410T-ZL, which are automatic watch-type wearable devices for self-BP measurement, according to the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-22013 guideline. Watches were held with the wrist at heart level. The mean differences between reference BPs and HEM-6410T-ZM readings were -0.9 ± 7.6/-1.1 ± 6.1 mm Hg for systolic BP (SBP)/diastolic BP (DBP) for criterion 1, and -0.9 ± 6.8/-1.1 ± 5.5 mm Hg for SBP/DBP for criterion 2. The mean differences between reference BPs and HEM-6410T-ZL readings were 2.4 ± 7.3/0.7 ± 7.0 mm Hg for SBP/DBP for criterion 1, and 2.4 ± 6.5/0.7 ± 6.5 mm Hg for SBP/DBP for criterion 2. The Omron HEM-6410T-ZM and the Omron HEM-6410T-ZL both fulfilled both validation criteria 1 and 2 of the ANSI/AAMI/ISO 81060-22013 guidelines.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Monitores de Presión Sanguínea / Monitoreo Ambulatorio de la Presión Arterial / Diseño de Equipo / Hipertensión Enmascarada / Dispositivos Electrónicos Vestibles Tipo de estudio: Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged País/Región como asunto: Asia Idioma: En Revista: J Clin Hypertens (Greenwich) Asunto de la revista: ANGIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Monitores de Presión Sanguínea / Monitoreo Ambulatorio de la Presión Arterial / Diseño de Equipo / Hipertensión Enmascarada / Dispositivos Electrónicos Vestibles Tipo de estudio: Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged País/Región como asunto: Asia Idioma: En Revista: J Clin Hypertens (Greenwich) Asunto de la revista: ANGIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Japón