Magnetic particle imaging for artifact-free imaging of intracranial flow diverter stents: A phantom study.
Herzberg, M; Dorn, F; Dietrich, P; Rückert, M A; Kampf, T; Bley, T A; Behr, V C; Herz, S; Vogel, P.
Afiliación
  • Herzberg M; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, 97080 Würzburg, Germany. Electronic address: Herzberg_M1@ukw.de.
  • Dorn F; Department of Neuroradiology, University Hospital Bonn, 53127 Bonn, Germany.
  • Dietrich P; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, 97080 Würzburg, Germany.
  • Rückert MA; Department of Experimental Physics 5 (Biophysics), University of Würzburg, 97074 Würzburg, Germany.
  • Kampf T; Department of Diagnostic and Interventional Neuroradiology, University Hospital Würzburg, 97080 Würzburg, Germany.
  • Bley TA; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, 97080 Würzburg, Germany.
  • Behr VC; Department of Experimental Physics 5 (Biophysics), University of Würzburg, 97074 Würzburg, Germany.
  • Herz S; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, 97080 Würzburg, Germany.
  • Vogel P; Department of Experimental Physics 5 (Biophysics), University of Würzburg, 97074 Würzburg, Germany.
Phys Med ; 88: 65-70, 2021 Aug.
Article en En | MEDLINE | ID: mdl-34192659
ABSTRACT

PURPOSE:

Magnetic Particle Imaging (MPI) is a new, background- and radiation-free tomographic imaging method that enables near real-time imaging of superparamagnetic iron-oxide nanoparticles (SPIONs) with high temporal and spatial resolution. This phantom study aims to investigate the potential of MPI for visualization of the stent lumen in intracranial flow diverters (FD).

METHODS:

Nitinol FD of different dimensions (outer diameter 3.5 mm, 4.0 mm, 5.5 mm; total length 22-40 mm) were scanned in vascular phantoms in a custom-built MPI scanner (in-plane resolution ~ 2 mm, field of view 65 mm length, 29 mm diameter). Phantoms were filled with diluted (150) SPION tracer agent Ferucarbotran (10 µmol (Fe)/ml; NaCL). Each phantom was measured in 32 different projections (overall acquisition time per image 3200 ms, 5averages). After image reconstruction from raw data, two radiologists assessed image quality using a 5-point Likert scale. The signal intensity profile was measured using a semi-automatic evaluation tool.

RESULTS:

MPI visualized the lumen of all FD without relevant differences between the stented vessel phantom and the reference phantom. At 3.5 mm image quality was slightly inferior to the larger diameters. The FD themselves neither generated an MPI signal nor did they lead to relevant imaging artifacts. Ratings of both radiologists showed no significant difference, interrater reliability was good (ICC 0.84). A quantitative evaluation of the signal intensity profile did not reveal any significant differences (p > 0.05) either.

CONCLUSION:

MPI visualizes the lumen of nitinol FD stents in vessel phantoms without relevant stent-induced artifacts.