A pumpless artificial lung without systemic anticoagulation: The Nitric Oxide Surface Anticoagulation system.
Fallon, Brian P; Lautner-Csorba, Orsolya; Thompson, Alex J; Lautner, Gergely; Kayden, Adrianna; Johnson, Matthew D; Harvey, Stephen L; Langley, Mark W; Peña, Alvaro Rojas; Bartlett, Robert H; Hirschl, Ronald B.
Afiliación
  • Fallon BP; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA. Electronic address: bfallon@med.umich.edu.
  • Lautner-Csorba O; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Thompson AJ; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Lautner G; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Kayden A; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Johnson MD; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Harvey SL; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Langley MW; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Peña AR; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Bartlett RH; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA.
  • Hirschl RB; Extracorporeal Life Support Laboratory, Department of Surgery, Michigan Medicine, B560 MSRB II/SPC 5686, 1150 W. Medical Center Drive, Ann Arbor, MI 48109, USA; Department of Surgery, Section of Pediatric Surgery, Michigan Medicine, Ann Arbor, MI, USA.
J Pediatr Surg ; 57(1): 26-33, 2022 Jan.
Article en En | MEDLINE | ID: mdl-34649727
ABSTRACT

BACKGROUND:

Artificial lungs have the potential to serve as a bridge to transplantation or recovery for children with end-stage lung disease dependent on extracorporeal life support, but such devices currently require systemic anticoagulation. We describe our experience using the novel Nitric Oxide (NO) Surface Anticoagulation (NOSA) system-an NO-releasing circuit with NO in the sweep gas-with the Pediatric MLung-a low-resistance, pumpless artificial lung.

METHODS:

NO flux testing MLungs (n = 4) were tested using veno-venous extracorporeal life support in a sheep under anesthesia with blood flow set to 0.5 and 1 L/min and sweep gas blended with 100 ppm NO at 1, 2, and 4 L/min. NO and NO2 were measured in the sweep and exhaust gas to calculate NO flux across the MLung membrane. Pumpless implants Sheep (20-100 kg, n = 3) underwent thoracotomy and cannulation via the pulmonary artery (device inflow) and left atrium (device outflow) using cannulae and circuit components coated with an NO donor (diazeniumdiolated dibutylhexanediamine; DBHD-N2O2) and argatroban. Animals were connected to the MLung with 100 ppm NO in the sweep gas under anesthesia for 24 h with no systemic anticoagulation after cannulation.

RESULTS:

NO flux testing NO flux averaged 3.4 ± 1.0 flux units (x10-10 mol/cm2/min) (human vascular endothelium 0.5-4 flux units). Pumpless implants 3 sheep survived 24 h with patent circuits. MLung blood flow was 716 ± 227 mL/min. Outlet oxygen saturation was 98.3 ± 2.6%. Activated clotting time was 151±24 s. Platelet count declined from 334,333 ± 112,225 to 123,667 ± 7,637 over 24 h. Plasma free hemoglobin and leukocyte and platelet activation did not significantly change.

CONCLUSIONS:

The NOSA system provides NO flux across a gas-exchange membrane of a pumpless artificial lung at a similar rate as native vascular endothelium and achieves effective local anticoagulation of an artificial lung circuit for 24 h.