Lead fixation mechanism impacts outcome of transvenous lead extraction: data from the European Lead Extraction ConTRolled Registry.
Levi, Nir; Bongiorni, Maria Grazia; Rav Acha, Moshe; Tovia-Brodie, Oholi; Kennergren, Charles; Auricchio, Angelo; Maggioni, Aldo Pietro; Rinaldi, Christopher A; Nof, Eyal; Ilan, Michael; Blomstrom-Lundqvist, Carina; Deharo, Jean-Claude; Leclercq, Christophe; Glikson, Michael; Michowitz, Yoav.
Afiliación
  • Levi N; Jesselson Integrated Heart Center, Shaare Zedek Medical Center, Faculty of Medicine Hebrew University, 12 Shmuel Beit Street, Jerusalem 9103102, Israel.
  • Bongiorni MG; Department of Cardiology, Azienda Ospedaliero-Universitaria, Pisa, Italy.
  • Rav Acha M; Jesselson Integrated Heart Center, Shaare Zedek Medical Center, Faculty of Medicine Hebrew University, 12 Shmuel Beit Street, Jerusalem 9103102, Israel.
  • Tovia-Brodie O; Jesselson Integrated Heart Center, Shaare Zedek Medical Center, Faculty of Medicine Hebrew University, 12 Shmuel Beit Street, Jerusalem 9103102, Israel.
  • Kennergren C; Department of Cardiothoracic Surgery, Sahlgrenska University Hospital, Gothenburg, Sweden.
  • Auricchio A; Division of Cardiology, Istituto Cardiocentro Ticino, Lugano, Switzerland.
  • Maggioni AP; ANMCO Research Center, Florence, Italy.
  • Rinaldi CA; Cardiology Department, St. Thomas' Hospital, London, UK.
  • Nof E; Leviev Heart Institute, Sheba Medical Center, Sackler School of Medicine, Tel Aviv University, Israel.
  • Ilan M; Jesselson Integrated Heart Center, Shaare Zedek Medical Center, Faculty of Medicine Hebrew University, 12 Shmuel Beit Street, Jerusalem 9103102, Israel.
  • Blomstrom-Lundqvist C; Department of Medical Science and Cardiology, Uppsala University, Uppsala, Sweden.
  • Deharo JC; Assistance Publique - Hôpitaux de Marseille, Centre Hospitalier Universitaire La Timone, Service de Cardiologie, Marseille, France.
  • Leclercq C; Aix Marseille University, C2VN Marseille, France.
  • Glikson M; Department of Cardiology, CHU Rennes, Rennes, France.
  • Michowitz Y; Jesselson Integrated Heart Center, Shaare Zedek Medical Center, Faculty of Medicine Hebrew University, 12 Shmuel Beit Street, Jerusalem 9103102, Israel.
Europace ; 24(5): 817-827, 2022 05 03.
Article en En | MEDLINE | ID: mdl-34652415
ABSTRACT

AIMS:

The aims of this study is to characterize the transvenous lead extraction (TLE) population with active (A) compared with passive fixation (PFix) leads and to compare the safety, efficacy, and ease of extracting active fixation (AFix) compared with PFix right atrial (RA) and right ventricular (RV) leads. METHODS AND

RESULTS:

The European Lead Extraction ConTRolled Registry (ELECTRa) was analysed. Patients were divided into three groups; those with only AFix, only PFix, and combined Fix leads. Three outcomes were defined. Difficult extraction, complete radiological, and clinical success. Multivariate model was used to analyse the independent effect of Fix mechanism on these outcomes. The study included 2815 patients, 1456 (51.7%) with only AFix leads, 982 (34.9%) with only PFix leads, and 377 (13.4%) with combined Fix leads. Patients with AFix leads were younger with shorter lead dwelling time. Infection was the leading cause for TLE among the combined Fix group with lowest rates among AFix group. No difference in complications rates was noted between patients with only AFix vs. PFix leads. Overall, there were 1689 RA (1046 AFix and 643 PFix) and 2617 RV leads (1441 AFix and 1176 PFix). Multivariate model demonstrated that PFix is independently associated with more difficult extraction for both RA and RV leads, lower radiological success in the RA but has no effect on clinical success.

CONCLUSION:

Mechanism of Fix impact the ease of TLE of RA and RV leads and rates of complete radiological success in the RA but not clinical success. These findings should be considered during implantation and TLE procedures.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Desfibriladores Implantables Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Europace Asunto de la revista: CARDIOLOGIA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Israel

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Desfibriladores Implantables Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Europace Asunto de la revista: CARDIOLOGIA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Israel