[Simulator of Heart Rate and Respiratory Rate for Monitoring Accuracy of Intelligent Sleep Monitoring Devices].
Fu, Li; Jiang, Shuo; Tian, Jia; Tan, Feifei; Chen, Tiantian; Zhang, Zaiai; Ding, Xiyan; Zhang, Ke.
Afiliación
  • Fu L; Shandong Institute of Medical Device and Pharmaceutical Packaging Inspection, Jinan, 250101.
  • Jiang S; NMPA Key Laboratory for Quality Evaluation of Medical Materials and Biological Protective Devices, Jinan, 250101.
  • Tian J; Shandong Institute of Medical Device and Pharmaceutical Packaging Inspection, Jinan, 250101.
  • Tan F; NMPA Key Laboratory for Quality Evaluation of Medical Materials and Biological Protective Devices, Jinan, 250101.
  • Chen T; Shandong Institute of Medical Device and Pharmaceutical Packaging Inspection, Jinan, 250101.
  • Zhang Z; NMPA Key Laboratory for Quality Evaluation of Medical Materials and Biological Protective Devices, Jinan, 250101.
  • Ding X; Shandong Institute of Medical Device and Pharmaceutical Packaging Inspection, Jinan, 250101.
  • Zhang K; NMPA Key Laboratory for Quality Evaluation of Medical Materials and Biological Protective Devices, Jinan, 250101.
Zhongguo Yi Liao Qi Xie Za Zhi ; 47(4): 445-448, 2023 Jul 30.
Article en Zh | MEDLINE | ID: mdl-37580299
ABSTRACT

OBJECTIVE:

Invent a simulator which provides a simulation of heart rate and respiratory rate to the intelligent sleep monitoring devices based on precision pressure sensors.

METHODS:

The simulator was composed of control part and simulated silicone doll. The simulated silicone doll contains heartbeat simulator and breathing simulation airbag. Heartbeat and breathing combination pressure signal can be produced according to frequency set values. Frequencies of pressure signal of the simulator were compared with the monitoring results of intelligent sleep monitoring devices with known accuracy to verify the frequency accuracy of pressure signal of the simulator. Verified the repeatability and stability of the simulator with a stopwatch.

RESULTS:

The heart rate of the simulator was with in ±2 beats per minute of the monitoring results of intelligent sleep monitoring devices and the respiratory rate of the simulator was with in ±2 times per minute of the monitoring results. The repeatability and stability of the simulator was better than ±5% according to results with a stopwatch.

CONCLUSIONS:

It's practicable to use the simulator which provides a simulation of heart rate and respiratory rate to the accuracy test of the intelligent sleep monitoring devices based on precision pressure sensor.