Performance of wearable watch-type home blood pressure measurement devices in a real-world clinical sample.
Lunardi, Mattia; Muhammad, Farooq; Shahzad, Atif; Nadeem, Asma; Combe, Lisa; Simpkin, Andrew J; Sharif, Faisal; Wijns, William; McEvoy, John W.
Afiliación
  • Lunardi M; Department of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and University of Galway, Galway, H91 TK33, Ireland.
  • Muhammad F; The Smart Sensors Laboratory at the Lambe Institute for Translational Medicine and CURAM, University of Galway, Galway, Ireland.
  • Shahzad A; Division of Cardiology, University Hospital of Verona, Verona, Italy.
  • Nadeem A; The Smart Sensors Laboratory at the Lambe Institute for Translational Medicine and CURAM, University of Galway, Galway, Ireland.
  • Combe L; The Smart Sensors Laboratory at the Lambe Institute for Translational Medicine and CURAM, University of Galway, Galway, Ireland.
  • Simpkin AJ; The Smart Sensors Laboratory at the Lambe Institute for Translational Medicine and CURAM, University of Galway, Galway, Ireland.
  • Sharif F; The Smart Sensors Laboratory at the Lambe Institute for Translational Medicine and CURAM, University of Galway, Galway, Ireland.
  • Wijns W; School of Mathematics, Statistics and Applied Mathematics, University of Galway, Galway, Ireland.
  • McEvoy JW; Insight Centre for Data Analytics, University of Galway, Galway, Ireland.
Clin Res Cardiol ; 113(10): 1393-1404, 2024 Oct.
Article en En | MEDLINE | ID: mdl-38112747
ABSTRACT

BACKGROUND:

Independent testing of home blood pressure (BP) measurement (HBPM) devices is often lacking, particularly among older and multi-morbid patients.

METHODS:

We studied the Bpro G2 (using tonometry), Omron HeartGuide (using occlusive oscillometric technology), and Heartisans (using photoplethysmography) wrist watch HBPM devices against a gold standard brachial sphygmomanometer. To test device performance, we used the ISO81060-2 protocol (though this protocol cannot formally validate cuffless devices). We also used linear mixed models to compare adjusted longitudinal BP measurements between devices. Finally, as a surrogate for usability, we recorded instances of device failure where no BP measurement was returned.

RESULTS:

We enrolled 128 participants (median [Q1-Q3] age 53 [40-65] years, 51% male, 46% on antihypertensive drugs), of whom 100 were suitable for the primary analysis. All three devices had mean BP values within 5 mmHg of sphygmomanometry. However, due to insufficient reliability (e.g., wider than accepted standard deviations of mean BP), none of the three devices passed all criteria required by the ISO81060-2 protocol. In adjusted longitudinal analyses, the Omron device also systematically underestimated systolic and diastolic BP (- 8.46 mmHg; 95% CI 6.07, 10.86; p < 0.001; and - 2.53 mmHg; 95% CI - 4.03, - 1.03; p = 0.001; respectively). Nevertheless, compared to the Omron device, BPro and Heartisans devices had increased odds of failure (BPro odds ratio [OR] 5.24; p < 0.0001; Heartisans OR 5.61; p < 0.001).

CONCLUSIONS:

While we could not formally validate the cuffless devices, our results show that wearable technologies will require improvements to offer reliable BP assessment. This study also highlights the need for validation protocols specifically designed for cuffless BP measurement technologies.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Monitoreo Ambulatorio de la Presión Arterial / Diseño de Equipo / Dispositivos Electrónicos Vestibles Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Res Cardiol Asunto de la revista: CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Irlanda

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Monitoreo Ambulatorio de la Presión Arterial / Diseño de Equipo / Dispositivos Electrónicos Vestibles Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Res Cardiol Asunto de la revista: CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Irlanda