Establishing a fingerprinting method for fast catheter identification in HDR brachytherapy in vivo dosimetry.
Tho, Daline; Bélanger, Cédric; Jørgensen, Erik B; Tanguay, Jérémie; Rosales, Haydee M L; Beddar, Sam; Johansen, Jacob G; Kertzscher, Gustavo; Lavallée, Marie-Claude; Beaulieu, Luc.
Afiliación
  • Tho D; Centre Intégré De Cancérologie, CHU De Québec, Université Laval, Centre De Recherche Chu De Québec, Québec, Canada; Département De Physique, De Génie Physique Et D'optique, Centre De Recherche Sur Le Cancer, Québec, Canada. Electronic address: daline.tho.1@ulaval.ca.
  • Bélanger C; Centre Intégré De Cancérologie, CHU De Québec, Université Laval, Centre De Recherche Chu De Québec, Québec, Canada; Département De Physique, De Génie Physique Et D'optique, Centre De Recherche Sur Le Cancer, Québec, Canada.
  • Jørgensen EB; Department of Oncology, Aarhus University Hospital, Aarhus, Denmark.
  • Tanguay J; Centre Intégré De Cancérologie, CHU De Québec, Université Laval, Centre De Recherche Chu De Québec, Québec, Canada; Département De Physique, De Génie Physique Et D'optique, Centre De Recherche Sur Le Cancer, Québec, Canada.
  • Rosales HML; Centre Intégré De Cancérologie, CHU De Québec, Université Laval, Centre De Recherche Chu De Québec, Québec, Canada; Département De Physique, De Génie Physique Et D'optique, Centre De Recherche Sur Le Cancer, Québec, Canada.
  • Beddar S; Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, TX.
  • Johansen JG; Department of Oncology, Aarhus University Hospital, Aarhus, Denmark.
  • Kertzscher G; Department of Oncology, Aarhus University Hospital, Aarhus, Denmark.
  • Lavallée MC; Centre Intégré De Cancérologie, CHU De Québec, Université Laval, Centre De Recherche Chu De Québec, Québec, Canada; Département De Physique, De Génie Physique Et D'optique, Centre De Recherche Sur Le Cancer, Québec, Canada.
  • Beaulieu L; Centre Intégré De Cancérologie, CHU De Québec, Université Laval, Centre De Recherche Chu De Québec, Québec, Canada; Département De Physique, De Génie Physique Et D'optique, Centre De Recherche Sur Le Cancer, Québec, Canada.
Brachytherapy ; 23(2): 165-172, 2024.
Article en En | MEDLINE | ID: mdl-38281894
ABSTRACT

PURPOSE:

To use quantities measurable during in vivo dosimetry to build unique channel identifiers, that enable detection of brachytherapy errors. MATERIALS AND

METHODS:

Treatment plan of 360 patients with prostate cancer who underwent high-dose-rate brachytherapy (range, 16-25 catheters; mean, 17) were used. A single point virtual dosimeter was placed at multiple positions within the treatment geometry, and the source-dosimeter distance and dwell time were determined for each dwell position in each catheter. These values were compared across all catheters, dwell position by dwell position, simulating a treatment delivery. A catheter was considered uniquely identified if, for a given dwell position, no other catheters had the same measured values. The minimum number of dwell positions needed to identify a specific catheter and the optimal dosimeter location uniquely were determined. The radial (r) and vertical (z) dimensions of the source-dosimeter distance were also examined for their utility in discriminating catheters.

RESULTS:

Using a virtual dosimeter with no uncertainties, all catheters were identified in 359 of the 360 cases with 9 dwell position measurements. When only the dwell time were measured, all catheters were uniquely identified after 1 dwell position. With a 2-mm spatial accuracy (r,z), all catheters were identified in 94% of the plans. Simultaneous measurement of source-dosimeter distance and dwell time ensured full catheter identification in all plans ranging from 2 to 6 dwell positions. The number of dwell positions needed to uniquely identify all catheters was lower when the distance from the implant center was higher.

CONCLUSIONS:

The most efficient fingerprinting approach involved combining source-dosimeter distance (i.e., source tracking) and dwell time. The further the dosimeter is placed from the center of the implant the better it can uniquely identify catheters.