Accuracy and Feasibility of Three-Dimensional Ultrasound Testing in Eye Clinic and Emergency Department Patients with Vision Complaints.
J Emerg Med
; 66(2): 197-210, 2024 02.
Article
en En
| MEDLINE
| ID: mdl-38309979
ABSTRACT
BACKGROUND:
Ocular emergencies comprise 2-3% of emergency department (ED) visits, with retinal detachment requiring emergency surgery. Two-dimensional ultrasound is a rapid bedside tool but is highly operator dependent.OBJECTIVE:
We determined three-dimensional ultrasound (3DUS) feasibility, acceptability, and usability in eye pathology detection using the ophthalmologist examination as reference standard.METHODS:
We performed a prospective, blinded cohort study of a 3DUS-enabling device in 30 eye clinic and ED patients with visual symptoms and calculated 3DUS performance characteristics. Two expert readers interpreted the 3DUS images for pathology. All participants completed surveys.RESULTS:
3DUS sensitivity was 0.81, specificity 0.73, positive predictive value 0.54, negative predictive value 0.91, and likelihood ratio (LR)+/LR- 3.03 and 0.26, respectively. Novice and expert sonographers had "substantial" agreement in correct diagnosis of abnormal vs. normal (κ = 0.68, 95% confidence interval 0.48-0.88). Most patients indicated that 3DUS is fast, comfortable, helps them understand their problem, and improves provider interaction/care, and all sonographers agreed; 4/5 sonographers felt confident performing ultrasound. Expert readers correctly identified an abnormal eye in 83/120 scans (76%) and correct diagnosis in 72/120 scans (65%), with no statistical difference between novice (79%; 69%) and expert (72%; 61%) sonographers (p = 0.39, p = 0.55), suggesting reduced operator dependence. Reader diagnosis confidence and image quality varied widely. Image acquisition times were fast for novice (mean 225 ± 83 s) and expert (201 ± 51) sonographers, with fast expert reader interpretation times (225 ± 136).CONCLUSIONS:
A 3DUS-enabling device demonstrates a sensitivity of 0.81 and specificity of 0.73 for disease detection, fast image acquisition, and may reduce operator dependence for detecting emergent retinal pathologies. Further technological development is needed to improve diagnostic accuracy in identifying and characterizing retinal pathology.Palabras clave
Texto completo:
1
Base de datos:
MEDLINE
Asunto principal:
Servicio de Urgencia en Hospital
Tipo de estudio:
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Límite:
Humans
Idioma:
En
Revista:
J Emerg Med
Asunto de la revista:
MEDICINA DE EMERGENCIA
Año:
2024
Tipo del documento:
Article