Your browser doesn't support javascript.
loading
Very Early Loading of a Bone-Anchored Hearing System Is Safe.
Møller, Marie Louise; Foghsgaard, Søren; Hougaard, Dan Dupont.
Afiliación
  • Møller ML; Department of Otorhinolaryngology, Head and Neck Surgery, Aalborg University Hospital, Aalborg, Denmark.
  • Foghsgaard S; Department of Otorhinolaryngology, Head and Neck Surgery, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark.
Otol Neurotol ; 45(5): 556-563, 2024 Jun 01.
Article en En | MEDLINE | ID: mdl-38693091
ABSTRACT

OBJECTIVE:

To investigate the feasibility of early sound processor loading with two separate bone-anchored hearing aid implants 1 to 2 weeks after implantation. Secondary endpoints included registration of any postoperative complications, skin reactions, and any adverse events. STUDY

DESIGN:

Prospective, nonrandomized, noncontrolled, multicenter clinical trial.

SETTING:

Tertiary referral center. PATIENTS Sixty adult patients, eligible for bone-anchored hearing aid surgery, were enrolled. INTERVENTION Implantation of the bone-anchored hearing aid implants were done using minimally invasive surgery with tissue preservation. MAIN OUTCOME

MEASURE:

Implant stability quotient (ISQ) values were recorded using resonance frequency analysis. Follow-ups were scheduled 5 to 12 days, 4 to 8 weeks, 6 months, and 12 months postoperatively. Differences between the lowest ISQ values (ISQ low) over time were analyzed by paired-samples t test with a significance level of 0.05. Skin and soft tissue reactions were assessed according to Holger's classification.

RESULTS:

Loading time of the Ponto Wide implant was 11.6 ± 1.4 days (SD) in average (median, 12 d; range, 7-19 d). Loading of the Ponto BHX implant was done 9.7 ± 3.5 days (SD) postoperatively (median, 9 d; range, 5-19 d). The implant stability increased significantly during the 12-month study period with both implants. One Ponto BHX implant was lost. Skin and soft tissue reactions were few in both groups. Across all planned follow-up visits, no skin and soft tissue reactions were observed in 86% (Ponto BHX implant) and 85% (Ponto Wide implant).

CONCLUSION:

Very early loading of sound processors after percutaneous bone-anchored hearing system surgery is safe already 1 to 2 weeks postoperatively.
Asunto(s)

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Audífonos Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Otol Neurotol Asunto de la revista: NEUROLOGIA / OTORRINOLARINGOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Dinamarca

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Audífonos Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Otol Neurotol Asunto de la revista: NEUROLOGIA / OTORRINOLARINGOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Dinamarca