Assessment by finite element analysis modelling of tissue strains associated with the use of two different nasogastric tube securement devices.
J Int Med Res
; 52(8): 3000605241264799, 2024 Aug.
Article
en En
| MEDLINE
| ID: mdl-39102503
ABSTRACT
OBJECTIVES:
Nasogastric tube use can lead to pressure injury. Some nasogastric tube securement devices (NG-SD) include hard plastic components. In the current study, we assessed the differences in strain profiles for two NG-SD, one with hard segments and one without hard segments, using finite element analysis (FEA) to measure strain and deformation occurring at the nasogastric tube-tissue interface.METHODS:
FEA in silico models of devices were based on device mechanical test data and clinically relevant placements. Peak strain values were determined by modelling different scenarios using Abaqus software whereby the tubing is moved during wear.RESULTS:
The modelling showed peak strains ranging from 52% to 434% for the two NG-SD depending on the tubing placement and device type. Peak strain was always higher for the hard plastic device. Tissue strain energy was a minimum of 133.8 mJ for the NG-SD with no hard parts and a maximum of 311.6 mJ for the NG-SD with hard parts.CONCLUSIONS:
This study provided evidence through in silico modelling that NG-SD without hard components may impart less strain and stress to tissues which may provide an option for tube securement that is less likely to cause medical device-related pressure injury.Palabras clave
Texto completo:
1
Base de datos:
MEDLINE
Asunto principal:
Simulación por Computador
/
Análisis de Elementos Finitos
/
Intubación Gastrointestinal
Límite:
Humans
Idioma:
En
Revista:
J Int Med Res
/
J. int. med. res
/
Journal of international medical research
Año:
2024
Tipo del documento:
Article
País de afiliación:
Estados Unidos