Your browser doesn't support javascript.
loading
Clinical outcomes and complications of a new high-density polyethylene-based spherical integrated porous orbital implant.
Garrido-Hermosilla, Antonio M; Díaz-Ruiz, María Concepción Concepción; Avilés-Prieto, Javier; Domínguez-Llamas, Sandra; Romero-Barranca, Isabel; Aguiar-Caro, José Antonio; Espejo-Arjona, Francisco; Relimpio-López, María Isabel.
Afiliación
  • Garrido-Hermosilla AM; Department of Ophthalmology, Virgen Macarena University Hospital, Seville, Spain. drgarridopublicaciones@gmail.com.
  • Díaz-Ruiz MCC; Retics OftaRed, Institute of Health Carlos III, Madrid, Spain. drgarridopublicaciones@gmail.com.
  • Avilés-Prieto J; Department of Ophthalmology, Virgen Macarena University Hospital, Seville, Spain.
  • Domínguez-Llamas S; Department of Ophthalmology, Virgen Macarena University Hospital, Seville, Spain.
  • Romero-Barranca I; Department of Ophthalmology, Virgen Macarena University Hospital, Seville, Spain.
  • Aguiar-Caro JA; Department of Ophthalmology, Virgen Macarena University Hospital, Seville, Spain.
  • Espejo-Arjona F; Department of Ophthalmology, Virgen Macarena University Hospital, Seville, Spain.
  • Relimpio-López MI; Department of Ophthalmology, Virgen Macarena University Hospital, Seville, Spain.
Int Ophthalmol ; 44(1): 371, 2024 Sep 06.
Article en En | MEDLINE | ID: mdl-39240285
ABSTRACT

PURPOSE:

To describe our experience over 6 years using a new high-density polyethylene-based spherical integrated porous orbital implants (Oculfit).

METHODS:

This is an observational retrospective case series study analyzing all cases requiring Oculfit implants between February 2015 and September 2021. Clinical information regarding the population included, the characteristics of the implant, and the outcomes and complications during the follow-up were noted. The success of the implant was defined according to anatomical and functional parameters.

RESULTS:

The study analyzed 90 cases of anophthalmic patients. The main causes for enucleation or evisceration were ocular decompensations (36.7%) and neoplasms (27.8% uveal melanoma and 7.8% retinoblastoma). Anatomical success was identified in 63 (70.0%) cases, functional success in 79 (87.8%) and complete success (anatomical + functional) in 61 (67.7%) cases. Factors associated with the functional success were age and exposure of the primary orbital implant. Complications appeared in 11 (12.2%) cases, which were completely resolved without sequelae in 4 (4.4%). Orbital explant was required in 5 (5.6%) cases.

CONCLUSION:

In our experience, Oculfit can be considered a useful alternative among the currently available options for orbital implants and has a good efficacy/safety profile.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Implantes Orbitales / Polietileno Límite: Adolescent / Adult / Aged / Aged80 / Child / Child, preschool / Female / Humans / Male / Middle aged Idioma: En Revista: Int Ophthalmol / Int. ophthalmol / International ophthalmology Año: 2024 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Implantes Orbitales / Polietileno Límite: Adolescent / Adult / Aged / Aged80 / Child / Child, preschool / Female / Humans / Male / Middle aged Idioma: En Revista: Int Ophthalmol / Int. ophthalmol / International ophthalmology Año: 2024 Tipo del documento: Article País de afiliación: España