RESUMO
BACKGROUND: An original methodology for determining the D antigen density on red cells was published in 2000 and has been applied in many publications since. This flow cytometry-based assay remained largely unrevised utilizing monoclonal anti-Ds that are not readily available anymore. We updated the methodology to quantify erythrocyte D antigen sites using microspheres and monoclonal anti-Ds that are commercially available today. METHODS: The absolute D antigen density of a frozen standard CcDEe cell, drawn in 2003, a fresh blood donation from the same individual, drawn in 2022, and an internal control CcDEe cell, was quantified by flow cytometry using fluorescence-labeled microspheres. The internal control CcDEe cell was used in conjunction with 9 commercial anti-Ds to determine D antigen densities of 7 normal D, 4 partial D, and 11 weak D type samples, including 2 novel alleles. RESULTS: The reproducibility of the updated assay was evaluated with red cells of published D antigen densities. The current results matched the known ones closely. The new weak D types 164 and 165 carried 4500 and 1505 D antigens/red cell, respectively. The absolute D antigen density decreased from 27,231 to 26,037 in an individual over 19 years. DISCUSSION: The updated assay gave highly reproducible results for the D antigen densities of Rh phenotypes. Readily available anti-Ds allowed for the determination of the D antigen densities of 7 weak D types. The assay is suitable to evaluate the effects of distinct amino acid substitutions on the RhD phenotype.