Purpose: To investigate the potential for decreasing radiation dose when utilizing a third-generation vs second-generation dual-source dual-energy CT (dsDECT) scanner, while maintaining diagnostic image quality and acceptable image noise. Materials and Methods: Retrospective analysis of patients who underwent dsDECT for clinical suspicion of urolithiasis from October 2, 2017, to September 5, 2018. Patient demographics, body mass index, abdominal diameter, scanning parameters, and CT dose index volume (CTDIvol) were recorded. Image quality was assessed by measuring the attenuation and standard deviation (SD) regions of interest in the aorta and in the bladder. Image noise was determined by averaging the SD at both levels. Patients were excluded if they had not undergone both third- and second-generation dual-energy CT (DECT), time between DECT was more than 2 years, or scan parameters were outside the standard protocol. Results: A total of 117 patients met the inclusion criteria. Examinations performed on a third-generation DECT had an average CTDIvol 12.3 mGy, while examinations performed on a second-generation DECT had an average CTDIvol 13.3 mGy (p < 0.001). Average image noise was significantly lower for the third-generation DECT (SD = 10.3) compared with the second-generation DECT (SD = 13.9) (p < 0.001). Conclusions: The third-generation dsDECT scanners can simultaneously decrease patient radiation dose and decrease image noise compared with second-generation DECT. These reductions in radiation exposure can be particularly important in patients with urinary stone disease who often require repeated imaging to evaluate for stone development and recurrence as well as treatment assessment.


Assuntos
Exposição à Radiação , Cálculos Urinários , Humanos , Doses de Radiação , Estudos Retrospectivos , Tomografia Computadorizada por Raios X , Cálculos Urinários/diagnóstico por imagem
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA