Resistencia a la corrosión y citotoxicidad de los aceros inoxidables expuestos a soluciones de cloruro / Corrosion resistance and cytotoxicity of stainless steels exposed to chloride solutions / Resistência à corrosão e citotoxicidade de aços inoxidáveis expostos a soluções de cloretos
Aguiar Júnior, Wagner de; Silva, Brunela Pereira da; Cruz, Aurea Silveira; Graziano, Kazuko Uchikawa; Aoki, Idalina Vieira.
Afiliação
  • Aguiar Júnior, Wagner de; Universidade de São Paulo, Escola Politécnica, Hospital Universitário. São Paulo. BR
  • Silva, Brunela Pereira da; Universidade de São Paulo, Escola Politécnica, Departamento de Engenharia Química. São Paulo. BR
  • Cruz, Aurea Silveira; Instituto Adolfo Lutz. São Paulo. BR
  • Graziano, Kazuko Uchikawa; Universidade de São Paulo, Escola de Enfermagem. São Paulo. BR
  • Aoki, Idalina Vieira; Universidade de São Paulo, Departamento de Engenharia Química. São Paulo. BR
Rev. SOBECC (Online) ; 29: E2429908, Fev. 2024. ilus
Article em En, Pt | LILACS | ID: biblio-1567245
Biblioteca responsável: BR2499
RESUMEN

Objetivo:

Analizar la resistencia a la corrosión por picaduras de aceros inoxidables AISI 304 y AISI 420 en un medio que contiene cloruros (solu-ción de NaCl al 0,9 y 3,5%, en masa), así como su citotoxicidad, in vitro, en muestras con y sin corrosión por picaduras.

Método:

Estudio experimental. Se utilizaron técnicas de polarización potenciodinámica cíclica (PPC) para caracterizar el alcance y la forma del ataque corrosivo a las muestras. Se utilizó el método de difusión en agar y evaluación de la viabilidad de la línea celular NCTC clon 929 (CCIAL 020) para evaluar la citotoxicidad de las muestras de acero con y sin picaduras.

Resultados:

El acero AISI 304 presentó una resistencia a la corrosión superior al acero AISI 420. Los valores de potencial de picadura disminuyeron para ambos aceros cuando aumentó la concentración de cloruros en la solución agresiva. Hubo toxicidad celular moderada (grado 3 ­ ISO 10993-5) en todas las muestras.

Conclusión:

Los resultados corroboraron las recomendaciones para evitar la inmersión innecesaria de instrumentos en soluciones salinas. La citotoxicidad moderada de estos aceros desaconseja su uso en dispositivos implantables, reservándolos solo para instrumentos quirúrgicos. (AU)
ABSTRACT

Objective:

To analyze the pitting corrosion resistance of AISI 304 and AISI 420 stainless steels in chloride-containing medium (0.9 and 3.5% NaCl solution, by weight), as well as their cytotoxicity,in vitro, in samples with and without pitting corrosion.

Method:

This is an experimental study. Cyclic potentiodynamic polarization (CPP) techniques were used to characterize the extent and shape of the corrosive attack on the samples. The agar diffusion and viability evaluation method of the NCTC clone 929 cell line (CCIAL 020) was used to evaluate the cytotoxicity of samples of steels with and without pitting.

Results:

The AISI 304 steel showed superior corrosion resistance to the AISI 420 steel. The values of the pitting potentials decreased for both steels when the chloride concentration in the aggressive solution was increased. There was moderate cell toxicity (grade 3 ­ ISO 10993-5) in all samples.

Conclusions:

The results corroborated the recommendations to avoid unnecessary immersion of the instruments in saline solutions. Moderate cytotoxicity to these steels contraindicates their use in implantable devices, only in surgical instruments. (AU)
RESUMO

Objetivo:

Analisar a resistência à corrosão por pites dos aços inoxidáveis AISI 304 e AISI 420 em meio contendo cloretos (solução de NaCl a 0,9 e 3,5%, em massa), assim como sua citotoxicidade,in vitro, em amostras com e sem corrosão por pites.

Método:

Estudo experimental. Utilizaram-se téc-nicas de polarização potenciodinâmica cíclica (PPC) para caracterizar extensão e forma do ataque corrosivo nas amostras. O método de difusão em ágar e avaliação da viabilidade da linhagem celular NCTC clone 929 (CCIAL 020) foi empregado para avaliar a citotoxicidade de amostras dos aços com e sem pites.

Resultados:

O aço AISI 304 apresentou resistência à corrosão superior ao aço AISI 420. Os valores dos potenciais de pite caíram para ambos os aços quando se aumentou a concentração de cloretos na solução agressiva. Houve moderada toxicidade celular (grau 3 ­ ISO 10993-5) em todas as amostras.

Conclusão:

Os resultados corroboraram as recomendações para evitar a imersão desnecessária dos instrumentais em soluções salinas. A citotoxicidade moderada para esses aços contraindica seu uso em dispositivos implantáveis, apenas em instrumentos cirúrgicos. (AU)