High dead volume pumps: delay time reduction by short-term low-ratio split flow in LC/MS applications.
Rogatsky, Eduard; Stein, Daniel T.
Afiliação
  • Rogatsky E; General Clinical Research Center, Albert Einstein College of Medicine of Yeshiva University, Bronx, New York 10461, USA. rogatsky@aecom.yu.edu
J Am Soc Mass Spectrom ; 18(2): 245-7, 2007 Feb.
Article em En | MEDLINE | ID: mdl-17095244
ABSTRACT
The aim of our work was to develop a low-cost, simple and reliable solution to reduce LC/MS analysis time by compensating for limitations inherent to high dead volume standard HPLC pumps. In our approach, we utilized a temporary (sub-1 min) low ratio flow split (approximately 110) at 5 mL/min pump flow before the column. During this short period, 90% of the entire pump flow is delivered to waste and used for fast pump/system equilibration. Although full-time flow splitting is widely used in capillary/nano applications (usually with high split ratios in the hundreds or thousands), to our knowledge, this is the first time that short-term low-ratio flow splitting has been used in conventional LC/MS applications.